FDA Enforcement Class II Ongoing

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Recall: Z-1024-2026 · Reported January 21, 2026

Enforcement

Recall Number
Z-1024-2026
Event ID
98260
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
REPLIGEN CORPORATION
Voluntary / Mandated
FDA Mandated
Report Date
January 21, 2026
Initiation Date
December 10, 2025
Classification Date
January 13, 2026
Address
111 Locke Dr, Marlborough, MA, 01752-7236, United States

Description

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Reason

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Code Info

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Distribution

U.S.

Quantity

33