FDA Enforcement Class II Terminated

SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system

Recall: Z-1024-2016 · Reported March 9, 2016

Enforcement

Recall Number
Z-1024-2016
Event ID
73333
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Electronics North America Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 9, 2016
Initiation Date
July 6, 2015
Classification Date
February 26, 2016
Termination Date
August 29, 2017
Address
3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States

Description

SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system

Reason

Wrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to instructions for use can lead to patient burns.

Code Info

4522 131 5575x 4522 132 1985x 4522 132 7619x 4598 000 5187x 4598 000 5189x 4598 003 59051

Distribution

US: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jersey Jordan Kazakhstan Kenya Korea, Republic of Kuwait Kyrgyzstan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Malta Martinique Mauritius Mexico Monaco Mongolia Morocco Mozambique Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian territory, occupied Panama Papua New Guinea Peru Philippines Poland Portugal Puerto Rico Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Venezuela Viet Nam Yemen Zimbabwe

Quantity

5672