FDA Enforcement Class II Terminated

20-Pole Eco Cable, Catalog #EM-5050-060 Product Usage: The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System.

Recall: Z-1023-2013 · Reported April 3, 2013

Enforcement

Recall Number
Z-1023-2013
Event ID
64579
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biosense Webster, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 3, 2013
Initiation Date
March 5, 2013
Classification Date
March 28, 2013
Termination Date
October 17, 2013
Address
15715 Arrow Hwy, Irwindale, CA, 91706-2006, United States

Description

20-Pole Eco Cable, Catalog #EM-5050-060 Product Usage: The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System.

Reason

Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field.

Code Info

Impacted serial Numbers: 120, 132, 134,136, 151, 342, 350, 354, 356, 363, 368, 373, 379, 380, 384, 385, 402, 405, 424, 919, 953, 954, 960, 963, 968, 970, 977, 980, 993, 998, 1001, 1123, 1148, 1378, 1596, 1729, 1745, 1747, 1749, 1750, 1751, 1753, 1754, 1758, 1759, 1760, 1761, 1797, 1799, 1802, 1803, 1804, 1805. Lot Numbers: P1210, P1220, P1221, P1231.

Distribution

Worldwide Distribution - USA (Nationwide) including the states of: DE, NL, CH, FR, GB, DK.

Quantity

54 units