FDA Enforcement
Class II
Terminated
Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
Recall: Z-1022-2013
·
Reported April 3, 2013
Enforcement
- Recall Number
- Z-1022-2013
- Event ID
- 64501
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Genetic Testing Institute,inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 3, 2013
- Initiation Date
- February 4, 2013
- Classification Date
- March 28, 2013
- Termination Date
- March 21, 2016
- Address
- 20925 Crossroads Cir Ste 200, Waukesha, WI, 53186-4054, United States
Description
Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA).
Reason
Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati
Code Info
Kit Lot Number 3000389
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
Quantity
386 kits (370 US, 16 OUS)