BrightView are gamma cameras; Designed for single or dual detector nuclear imaging.
Enforcement
- Recall Number
- Z-1020-2016
- Event ID
- 73010
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 9, 2016
- Initiation Date
- December 17, 2015
- Classification Date
- February 26, 2016
- Termination Date
- July 2, 2018
- Address
- 595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States
Description
BrightView are gamma cameras; Designed for single or dual detector nuclear imaging.
The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan. Issue 2: During a scan, the gantry motion may halt prematurely. This cause
Model # 882480; Serial #'s: 4000002, 4000002.75, 4000003, 4000004, 4000006, 4000007, 4000008, 4000009, 4000010, 4000011, 4000013, 4000030, 4000031, 4000032, 4000033, 4000034, 4000035, 4000036, 4000037, 4000038, 4000039, 4000040, 4000041, 4000043, 4000044, 4000045, 4000046, 4000047, 4000048, 4000049, 4000050, 4000051, 4000052, 4000053, 4000054, 4000055, 4000056, 4000057, 4000058, 4000059, 4000060, 4000061, 4000062, 4000064, 4000065, 4000066, 4000068, 4000070, 4000071, 4000072, 4000073, 4000074, 4000075, 4000076, 4000077, 4000078, 4000079, 4000080, 4000081, 4000082, 4000083, 4000086, 4000087, 4000088, 4000089, 4000090, 4000091, 4000093, 4000094, 4000095, 4000096, 4000099, 4000100, 4000102, 4000103, 4000104, 4000105, 4000106, 4000107, 4000108, 4000109, 4000110, 4000111, 4000112, 4000113, 4000114, 4000115, 4000116, 4000117, 4000118, 4000119, 4000120, 4000121, 4000122, 4000123, 4000124, 4000125, 4000126, 4000127, 4000128, 4000129, 4000130, 4000132, 4000133, 4000134, 4000135, 4000136, 4000137, 4000138, 4000139, 4000140, 4000141, 4000142, 4000143, 4000144, 4000145, 4000146, 4000148, 4000149, 4000151, 4000152, 4000153, 4000154, 4000155, 4000156, 4000157, 4000158, 4000159, 4000160, 4000161, 4000162, 4000163, 4000164, 4000165, 4000166, 4000167, 4000168, 4000169, 4000170, 4000171, 4000173, 4000174, 4000175, 4000176, 4000177, 4000178, 4000179, 4000180, 4000181, 4000182, 4000183, 4000184, 4000185, 4000186, 4000187, 4000188, 4000189, 4000190, 4000191, 4000192, 4000193, 4000194, 4000196, 4000197, 4000198, 4000199, 4000200, 4000202, 4000203, 4000204, 4000205, 4000206, 4000207, 4000208, 4000209, 4000211, 4000212, 4000214, 4000215, 4000216, 4000218, 4000219, 4000220, 4000221, 4000222, 4000223, 4000224, 4000225, 4000226, 4000227, 4000228, 4000229, 4000230, 4000231, 4000232, 4000233, 4000234, 4000235, 4000236, 4000237, 4000238, 4000239, 4000240, 4000241, 4000242, 4000243, 4000244, 4000245, 4000246, 4000247, 4000248, 4000249, 4000250, 4000251, 4000253, 4000254, 4000255, 4000256, 4000257, 4000258, 4000259, 4000260, 4000261, 4000262, 4000263, 4000264, 4000265, 4000266, 4000267, 4000268, 4000269, 4000270, 4000271, 4000272, 4000273, 4000274, 4000275, 4000276, 4000277, 4000278, 4000279, 4000280, 4000281, 4000282, 4000283, 4000284, 4000285, 4000286, 4000287, 4000288, 4000289, 4000290, 4000291, 4000292, 4000293, 4000294, 4000295, 4000296, 4000297, 4000298, 4000299, 4000300, 4000301, 4000302, 4000303, 4000304, 4000305, 4000306, 4000307, 4000308, 4000309, 4000310, 4000311, 4000312, 4000313, 4000314, 4000315, 4000316, 4000317, 4000318, 4000319, 4000320, 4000321, 4000322, 4000323, 4000324, 4000325, 4000326, 4000327, 4000328, 4000329, 4000330, 4000331, 4000332, 4000333, 4000334, 4000335, 4000336, 4000337, 4000338, 4000339, 4000340, 4000341, 4000342, 4000343, 4000344, 4000345, 4000346, 4000347, 4000348, 4000349, 4000351, 4000352, 4000353, 4000354, 4000355, 4000356, 4000357, 4000358, 4000359, 4000360, 4000361, 4000362, 4000363, 4000364, 4000366, 4000367, 4000368, 4000369, 4000370, 4000371, 4000372, 4000373, 4000374, 4000375, 4000376, 4000377, 4000378, 4000379, 4000380, 4000381, 4000382, 4000383, 4000384, 4000385, 4000386, 4000387, 4000388, 4000389, 4000390, 4000391, 4000392, 4000394, 4000395, 4000396, 4000397, 4000398, 4000399, 4000400, 4000401, 4000402, 4000403, 4000404, 4000405, 4000406, 4000407, 4000408, 4000409, 4000410, 4000411, 4000412, 4000414, 4000415, 4000416, 4000417, 4000418, 4000419, 4000420, 4000421, 4000422, 4000423, 4000424, 4000425, 4000426, 4000428, 4000429, 4000430, 4000431, 4000432, 4000433, 4000434, 4000435, 4000436, 4000437, 4000438, 4000439, 4000440, 4000441, 4000442, 4000443, 4000444, 4000445, 4000446, 4000447, 4000448, 4000449, 4000450, 4000451, 4000452, 4000453, 4000454, 4000455, 4000456, 4000459, 4000460, 4000461, 4000462, 4000463, 4000464, 4000465, 4000466, 4000467, 4000468, 4000469, 4000470, 4000471, 4000473, 4000474, 4000475, 4000476, 4000477, 4000478, 4000479, 4000480, 4000481, 4000482, 4000483, 4000484, 4000485, 4000486, 4000487, 4000488, 4000489, 4000490, 4000491, 4000492, 4000493, 4000494, 4000495, 4000496, 4000497, 4000498, 4000499, 4000500, 4000501, 4000502, 4000503, 4000504, 4000505, 4000506, 4000507, 4000510, 4000511, 4000513, 4000516, 4000517, 4000518, 4000519, 4000520, 4000521, 4000522, 4000523, 4000524, 4000525, 4000526, 4000527, 4000528, 4000529, 4000530, 4000531, 4000532, 4000533, 4000534, 4000535, 4000536, 4000537, 4000538, 4000539, 4000540, 4000541, 4000542, 4000543, 4000544, 4000545, 4000546, 4000547, 4000549, 4000550, 4000551, 4000552, 4000553, 4000554, 4000555, 4000556, 4000557, 4000558, 4000559, 4000560, 4000561, 4000562, 4000563, 4000564, 4000565, 4000566, 4000567, 4000568, 4000569, 4000571, 4000572, 4000573, 4000574, 4000575, 4000576, 4000577, 4000578, 4000580, 4000581, 4000582, 4000583, 4000584, 4000585, 4000586, 4000587, 4000588, 4000589, 4000590, 4000591, 4000592, 4000593, 4000594, 4000595, 4000596, 4000597, 4000598, 4000599, 4000600, 4000601, 4000602, 4000603, 4000604, 4000605, 4000606, 4000607, 4000608, 4000609, 4000610, 4000611, 4000612, 4000613, 4000614, 4000615, 4000616, 4000617, 4000618, 4000619, 4000620, 4000621, 4000622, 4000623, 4000624, 4000625, 4000626, 4000627, 4000628, 4000629, 4000630, 4000631, 4000632, 4000633, 4000634, 4000635, 4000636, 4000637, 4000638, 4000639, 4000640, 4000641, 4000642, 4000643, 4000644, 4000645, 4000646, 4000647, 4000648, 4000649, 4000650, 4000651, 4000652, 4000653, 4000654, 4000655, 4000656, 4000657, 4000658, 4000659, 4000660, 4000661, 4000662, 4000663, 4000664, 4000665, 4000666, 4000667, 4000668, 4000669, 4000670, 4000671, 4000672, 4000673, 4000674, 4000675, 4000676, 4000677, 4000678, 4000679, 4000680, 4000681, 4000682, 4000683, 4000684, 4000685, 4000686, 4000687, 4000688, 4000689, 4000690, 4000691, 4000692, 4000693, 4000694, 4000695, 4000696, 4000697, 4000698, 4000699, 4340004, 4340005, 40000092, 40000101, 4000703001, 40007030003, 40007050005, 40007060008, 4000057-OLD, 4000098A, 4000151-1, 4000301-1, 4000472A, 4000484A, 10422, 400054, 4000029, 4000063, 4000067, 4000150, 4000195, 4000201, 4000210, 4000217, 4000252, 4000508, 4000509, 40007060006, 40007060007, 4000012, 4000213 & 4000413.
Worldwide Distribution-US (nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WA, WI, WV & WY, and in DC and PR and in the countries of Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Cayman Islands, China, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Gabon, Germany, Greece, India, Indonesia, Iran, Ireland, Italy, Japan, Korea, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Palestinian, Panama, Philippines, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom & Viet Nam.
677 units