FDA Enforcement Class II Terminated

aap Implante AG, Lambotte chisel, 20 mm, REF IU 3000-20 Product Usage: Surgical instruments are intended to aid in the insertion of bone plates and screws as well as for reduction and support during orthopedic surgeries. If not otherwise specified, instruments are intended for multiple use.

Recall: Z-1019-2019 · Reported March 27, 2019

Enforcement

Recall Number
Z-1019-2019
Event ID
82159
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
AAP Implantate Ag
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 27, 2019
Initiation Date
January 11, 2019
Classification Date
March 17, 2019
Termination Date
May 26, 2022
Address
Lorenzweg 5, N/A, Berlin, N/A, N/A, Germany

Description

aap Implante AG, Lambotte chisel, 20 mm, REF IU 3000-20 Product Usage: Surgical instruments are intended to aid in the insertion of bone plates and screws as well as for reduction and support during orthopedic surgeries. If not otherwise specified, instruments are intended for multiple use.

Reason

There is the potential for the Lambotte chisel to break at the welded piece parts of the chisel body and knocking plate.

Code Info

Lot Numbers: I011, I012, I013, I014, I015, I016, I017

Distribution

Worldwide Distribution - US Nationwide in the state of Texas and country of Germany

Quantity

25 units