FDA Enforcement
Class II
Terminated
BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10
Recall: Z-1018-2022
·
Reported May 11, 2022
Enforcement
- Recall Number
- Z-1018-2022
- Event ID
- 89896
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 11, 2022
- Initiation Date
- March 9, 2022
- Classification Date
- April 29, 2022
- Termination Date
- January 19, 2024
- Address
- 7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States
Description
BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10
Reason
When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.
Code Info
All systems running BD Synapsys version 4.10; UDI: 00382904441500; Serial Numbers: SL00530 SL00515 SL00527 SL00502 SL00540 SL00564 SL00549 SL00509 SL00003 SL00004 SL00010 SL00001 SL00005 SL00002 SL000015 SL01010 SL01022
Distribution
Worldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.
Quantity
15 systems