FDA Enforcement
Class II
Ongoing
Fructosamine Control 1, FR2994 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.
Recall: Z-1017-2019
·
Reported March 27, 2019
Enforcement
- Recall Number
- Z-1017-2019
- Event ID
- 81783
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Randox Laboratories Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 27, 2019
- Initiation Date
- December 3, 2018
- Classification Date
- March 15, 2019
- Address
- 55 Diamond Road, N/A, Crumlin (North), N/A, N/A, Ireland
Description
Fructosamine Control 1, FR2994 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.
Reason
The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.
Code Info
499FR
Distribution
US Distribution to states to: CA, ME. MI, and WV.
Quantity
141 kits