FDA Enforcement Class II Ongoing

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097

Recall: Z-1014-2022 · Reported May 4, 2022

Enforcement

Recall Number
Z-1014-2022
Event ID
89894
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Instrumentation Laboratory
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 4, 2022
Initiation Date
February 22, 2022
Classification Date
April 28, 2022
Address
180 Hartwell Rd, Bedford, MA, 01730-2443, United States

Description

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097

Reason

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Code Info

All Serial Numbers. Trade Name Part Number UDI ACL TOP 300 CTS 00000280060 08426950556916; "ACL TOP 300 CTS (Refurbished)" 00000280060R 08426950928027; ACL TOP 350 CTS 00000280065 08426950784081; ACL TOP 500 CTS 00000280040 08426950453499; "ACL TOP 500 CTS (Refurbished)" 00000280040R 08426950928003; ACL TOP 550 CTS 00000280045 08426950729242; ACL TOP 970 CL 00000280097 08430793045476.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, El Salvador, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, S. Tome, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad, Tobago, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam.

Quantity

2725 units US: 5,572 units OUS.UPDATE: 2727 units US; 5277 units OUS