FDA Enforcement Class II Terminated

NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Recall: Z-1010-2018 · Reported March 21, 2018

Enforcement

Recall Number
Z-1010-2018
Event ID
79236
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fresenius Medical Care Renal Therapies Group, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
March 21, 2018
Initiation Date
September 29, 2017
Classification Date
March 14, 2018
Termination Date
August 9, 2018
Address
920 Winter St, Waltham, MA, 02451-1521, United States

Description

NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason

The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.

Code Info

Lot # 17KTAC076; UDI-00840861101696

Distribution

US Distribution to states of: CA, FL, KS, LA, and TX.

Quantity

598 cases