symmetry surgical ULTRA(R) Steam container, REF numbers: C131310, C13134, C13135, C13136, C13139, C13854, C13855, C13856, C181310, C18134, C18135, C18136, C18138, C22854, C22855, C22856, C22858, C24134, C24135, C24136 Product Usage: sterilization container
Recall: Z-1010-2017
·
Reported January 18, 2017
Enforcement
- Recall Number
- Z-1010-2017
- Event ID
- 75850
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Symmetry Surgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 18, 2017
- Initiation Date
- November 29, 2016
- Classification Date
- January 11, 2017
- Termination Date
- August 20, 2017
- Address
- 3034 Owen Dr, Antioch, TN, 37013-2413, United States
Description
symmetry surgical ULTRA(R) Steam container, REF numbers: C131310, C13134, C13135, C13136, C13139, C13854, C13855, C13856, C181310, C18134, C18135, C18136, C18138, C22854, C22855, C22856, C22858, C24134, C24135, C24136 Product Usage: sterilization container
Reason
Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
Code Info
All
Distribution
Worldwide Distribution
Quantity
7613 units