FDA Enforcement Class II Ongoing

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Recall: Z-1008-2026 · Reported January 7, 2026

Enforcement

Recall Number
Z-1008-2026
Event ID
98076
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beaver-Visitec International, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 7, 2026
Initiation Date
December 5, 2025
Classification Date
December 31, 2025
Address
500 Totten Pond Rd, N/A, Waltham, MA, 02451-1916, United States

Description

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Reason

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

Code Info

Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

Distribution

US Nationwide distribution in the states of FL, MI, OK, TX, WI.

Quantity

5 units