FDA Enforcement
Class I
Terminated
MAQUET Servo Humidifier 163; Model No.: 01-06-8125-8, catalog no. XKC01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.
Recall: Z-1008-2015
·
Reported February 18, 2015
Enforcement
- Recall Number
- Z-1008-2015
- Event ID
- 70127
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Teleflex, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 18, 2015
- Initiation Date
- December 4, 2014
- Classification Date
- February 6, 2015
- Termination Date
- March 21, 2017
- Address
- 550 E Swedesford Rd Ste 400, N/A, Wayne, PA, 19087-1601, United States
Description
MAQUET Servo Humidifier 163; Model No.: 01-06-8125-8, catalog no. XKC01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.
Reason
Connector mount cracks which may lead to a leak failure during use and necessitate immediate replacement in the breathing circuit. If leak exists and is left untreated, patient may be deprived of adequate anesthetic gases over a period of time and serious adverse health consequences may occur.
Code Info
Batch Numbers: 201321, 201323, and 201324.
Distribution
Distributed in the states of FL, MI, WV, and KS.
Quantity
18