FDA Enforcement Class II Terminated

Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011

Recall: Z-1001-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-1001-2020
Event ID
84614
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Clinical Diagnostic Solutions, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 12, 2020
Initiation Date
January 2, 2020
Classification Date
February 5, 2020
Termination Date
November 3, 2022
Address
1800 NW 65th Ave, N/A, Plantation, FL, 33313-4544, United States

Description

Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011

Reason

A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.

Code Info

Part Number 1400075 (US) Part Number 1400011 (OUS) All serial numbers distributed prior to 3/21/2019

Distribution

Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.

Quantity

261