FDA Enforcement Class III Terminated

R&D CBC-3D Hematology Control: (1) CBC-3D Normal Pack Hematology Control, REF 3D501, containing 10 x 2.0mL Control N; (2) CBC-3D TRI PACK Hematology Control, REF 3D503, containing 4 x 2.0 mL Controls L, N, and H; (3) CBC-3D for Mythic 18 Hematology Control, REF 3D503RX, containing 2 x 2.0 mL each Controls L, N, and H; (4) CBC-3D For MINDRAY Hematology Control, REF 3D506, containing 2 x 3.0 mL each Controls L, N, and H; (5) CBC-3D For MINDRAY Hematology Control, REF 3D506US, containing 2 x 3.0 mL each Controls L, N, and H; (6) CBC-3D TRI PACK Hematology Control, REF 3D507, containing 4 x 2.0 mL each Controls L, N, and H; (7) CBC-3D TRI PACK Hematology Control, REF 3D508, containing 2 x 2.0 m each Controls L, N, and H; (8) CBC-3D For ABACUS, REF D3D02 Hematology Control, containing 2 x 3.0 mL each Controls L, N, and H; (9) CBC-3D For ABACUS Hematology Control, REF D3D04, containing 4 x 3.0 mL each Controls L, N, and H. Loose tubes labeled CBC-3D Hematology Control in 2.0mL and 3.0mL sizes. OEM/Own-label distributor labels: (1) Alere 380 Hemo Control (L-N-H), REF AL35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H; Kit lot #B1122; control lot #B1122N, no UDI (2) diatron Diacon 3 Hematology Control, Normal Level, 3.0 mL tubes, lot #B1122N (3) diatron Diacon 3 Hematology Control, REF DN35002-SET, containing 6 x 3.0 mL Control N, kit lot #B1122, control lot #B1122N (4) diatron Diacon 3 Hematology Control, REF DN35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, Normal control lot #B1122N. (5) Elite 3 HEM Control Normal, REF HEM00014, containing 6 x 3.0 mL Normal controls, kit lot #B1122, control lot #B1122N; (6) Erba H3 CON N, REF HEM00034, 3 mL vials, lot #E1122N, exp. 2023-02-10 (7) Horiba Medical Minotrol 16 Whole Blood Hematology Control, Cat. #5300000277, containing 3 x 2.5mL tubes (1 each Control 1, Control 2 (Normal), and Control 3; Cat. #5300000278, containing 6 x 2.5mL tubes (2 each Control 1, Control 2 (Normal), and Control 3; and Cat. #5300000283, containing 12 x 2.5mL tubes (4 each Control 1, Control 2 (Normal), and Control 3; kit lot #MX438, exp. 2023-01-05, Control 2 lot #MX438N. (8) 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-912-3D; and 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-914-3D. (9) mindray BC-3D, P/N 040-001423-00. (10) MINDRAY BC-3D CONTROL, REF BC3D01, containing 1 x 3.0 mL Controls L, N, and H; REF BC3D02, containing 2 x 3.0 mL Controls L, N, and H; and REF BC3D6N, containing 6 x 3.0 mL Control N; kit lot #B1122, control lot #B1122N. (11) MINDRAY BC-3D CONTROL, REF BC3D02US, containing 2 x 3.0 mL Level L, N, and H, kit lot number B1122, control lot #B1122N. (12) Nihon Kohden MEK-3D Hematology Control, REF 3DN03, containing MEK-3DN x 3 tubes; REF 3DN06, containing MEK-3DN x 6 tubes; and REF 3DN12, containing MEK-3DN x 12 tubes, kit lot #B22Y, control lot #B22YN. (13) Nihon Kohden MEK-3DN Normal Hematology Control tubes, 2.0 mL, lot #B22YN. (14) Orphee MYT-3D, REF MYT301, containing 2 x 2.5 mL Control N; REF MYT302, containing 2 x 2.5 mL each Controls L, N, and H; REF MYT304N, containing 4 x 2.5 mL Control N; REF 306N, containing 6 x 2.5 mL Control N; kit lot #B1122, control lot #B1122N; Myt-3D VET, REF MYT303-VET, containing 1 x 2.5 mL each Control L, N, and H. (15) Siemens ADVIA 360 Control, REF 11170853, containing 2 x 3.0 mL each Controls Low, Normal & High, kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) URIT HQ-3DIFF, REF 5411000306, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, control lot #B1122N; URIT HQ-3DIFF Hematology Control N tubes, 3.0 mL, lot #B1122N. (17) Woodley Haematology Control, REF WD1154, containing 6 x 2.0 mL Control N; and REF WD1154A, containing 2 x 2.0 mL each Controls L, N, and H, kit lot number B1122, control lot #B1122N;

Recall: Z-0999-2023 · Reported January 25, 2023

Enforcement

Recall Number
Z-0999-2023
Event ID
91136
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
R & D Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 25, 2023
Initiation Date
November 7, 2022
Classification Date
January 19, 2023
Termination Date
May 26, 2026
Address
614 Mckinley Pl Ne, N/A, Minneapolis, MN, 55413-2610, United States

Description

R&D CBC-3D Hematology Control: (1) CBC-3D Normal Pack Hematology Control, REF 3D501, containing 10 x 2.0mL Control N; (2) CBC-3D TRI PACK Hematology Control, REF 3D503, containing 4 x 2.0 mL Controls L, N, and H; (3) CBC-3D for Mythic 18 Hematology Control, REF 3D503RX, containing 2 x 2.0 mL each Controls L, N, and H; (4) CBC-3D For MINDRAY Hematology Control, REF 3D506, containing 2 x 3.0 mL each Controls L, N, and H; (5) CBC-3D For MINDRAY Hematology Control, REF 3D506US, containing 2 x 3.0 mL each Controls L, N, and H; (6) CBC-3D TRI PACK Hematology Control, REF 3D507, containing 4 x 2.0 mL each Controls L, N, and H; (7) CBC-3D TRI PACK Hematology Control, REF 3D508, containing 2 x 2.0 m each Controls L, N, and H; (8) CBC-3D For ABACUS, REF D3D02 Hematology Control, containing 2 x 3.0 mL each Controls L, N, and H; (9) CBC-3D For ABACUS Hematology Control, REF D3D04, containing 4 x 3.0 mL each Controls L, N, and H. Loose tubes labeled CBC-3D Hematology Control in 2.0mL and 3.0mL sizes. OEM/Own-label distributor labels: (1) Alere 380 Hemo Control (L-N-H), REF AL35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H; Kit lot #B1122; control lot #B1122N, no UDI (2) diatron Diacon 3 Hematology Control, Normal Level, 3.0 mL tubes, lot #B1122N (3) diatron Diacon 3 Hematology Control, REF DN35002-SET, containing 6 x 3.0 mL Control N, kit lot #B1122, control lot #B1122N (4) diatron Diacon 3 Hematology Control, REF DN35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, Normal control lot #B1122N. (5) Elite 3 HEM Control Normal, REF HEM00014, containing 6 x 3.0 mL Normal controls, kit lot #B1122, control lot #B1122N; (6) Erba H3 CON N, REF HEM00034, 3 mL vials, lot #E1122N, exp. 2023-02-10 (7) Horiba Medical Minotrol 16 Whole Blood Hematology Control, Cat. #5300000277, containing 3 x 2.5mL tubes (1 each Control 1, Control 2 (Normal), and Control 3; Cat. #5300000278, containing 6 x 2.5mL tubes (2 each Control 1, Control 2 (Normal), and Control 3; and Cat. #5300000283, containing 12 x 2.5mL tubes (4 each Control 1, Control 2 (Normal), and Control 3; kit lot #MX438, exp. 2023-01-05, Control 2 lot #MX438N. (8) 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-912-3D; and 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-914-3D. (9) mindray BC-3D, P/N 040-001423-00. (10) MINDRAY BC-3D CONTROL, REF BC3D01, containing 1 x 3.0 mL Controls L, N, and H; REF BC3D02, containing 2 x 3.0 mL Controls L, N, and H; and REF BC3D6N, containing 6 x 3.0 mL Control N; kit lot #B1122, control lot #B1122N. (11) MINDRAY BC-3D CONTROL, REF BC3D02US, containing 2 x 3.0 mL Level L, N, and H, kit lot number B1122, control lot #B1122N. (12) Nihon Kohden MEK-3D Hematology Control, REF 3DN03, containing MEK-3DN x 3 tubes; REF 3DN06, containing MEK-3DN x 6 tubes; and REF 3DN12, containing MEK-3DN x 12 tubes, kit lot #B22Y, control lot #B22YN. (13) Nihon Kohden MEK-3DN Normal Hematology Control tubes, 2.0 mL, lot #B22YN. (14) Orphee MYT-3D, REF MYT301, containing 2 x 2.5 mL Control N; REF MYT302, containing 2 x 2.5 mL each Controls L, N, and H; REF MYT304N, containing 4 x 2.5 mL Control N; REF 306N, containing 6 x 2.5 mL Control N; kit lot #B1122, control lot #B1122N; Myt-3D VET, REF MYT303-VET, containing 1 x 2.5 mL each Control L, N, and H. (15) Siemens ADVIA 360 Control, REF 11170853, containing 2 x 3.0 mL each Controls Low, Normal & High, kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) URIT HQ-3DIFF, REF 5411000306, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, control lot #B1122N; URIT HQ-3DIFF Hematology Control N tubes, 3.0 mL, lot #B1122N. (17) Woodley Haematology Control, REF WD1154, containing 6 x 2.0 mL Control N; and REF WD1154A, containing 2 x 2.0 mL each Controls L, N, and H, kit lot number B1122, control lot #B1122N;

Reason

For one lot of CBC-3D Hematology Control, the Normal Level may exhibit hemolysis or deterioration due to microbial contamination.

Code Info

R&D labeled product lot numbers: (Products 1-9): Lot B1122, exp. date 2023-02-05; UDI numbers for the following catalog numbers: 3D501 - UDI 10815762020091; 3D503 - UDI 00815762020117; 3D503RX - UDI 00815762020124; 3D506 - UDI N/A OUS only; 3D506US - UDI 00815762020162; 3D507 - UDI 00815762020179; 3D508 - UDI 00815762020186; D3D02 - UDI 00815762020230; and D3D04 - UDI 00815762020247. OEM/Own-label distributor labeled product lot numbers (Products 1-17) (1) Kit lot #B1122; control lot #B1122N. (2) Lot #B1122N. (3) Kit lot #B1122, control lot #B1122N. (4) Kit lot #B1122, control lot #B1122N. (5) Kit lot #B1122, control lot #B1122N. (6) Lot #E1122N. (7) Kit lot #MX438, Control 2 lot #MX438N. (8) Lot #2022-912-3D and Lot #2022-914-3D. (9) P/N 040-001423-00. (10) Kit lot #B1122, control lot #B1122N. (11) Kit lot number B1122, control lot #B1122N. (12) Kit lot #B22Y, control lot #B22YN. (13) Lot #B22YN. (14) Kit lot #B1122, control lot #B1122N. (15) Kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) Kit lot #B1122, control lot #B1122N. (17) Kit lot number B1122, control lot #B1122N;

Distribution

Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IA, KY, LA, MA, MO, NC, NJ, NY, OH, TX and Puerto Rico. The countries of Brazil, China, Costa Rica, Czech Republic, France, Germany, Great Britain, Honduras, Hungary, Indonesia, Israel, Japan, Kuwait, Mexico, Netherlands, Nicaragua, Panama, Peru, Poland, Romania, San Salvador, South Africa, South Korea, Switzerland, Thailand, Turkey, Vietnam,.

Quantity

8,259 kits, 28,136 labeled loose tubes, and 15,926 unlabeled loose tubes