FDA Enforcement Class II Terminated

Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK

Recall: Z-0997-2019 · Reported March 20, 2019

Enforcement

Recall Number
Z-0997-2019
Event ID
80520
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ossur Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 20, 2019
Initiation Date
June 21, 2018
Classification Date
March 11, 2019
Termination Date
September 4, 2020
Address
27051 Towne Centre Dr, N/A, Foothill Ranch, CA, 92610-2804, United States

Description

Ossur Skull Pin, Sterile Titanium Pin, Part/Description: 516TS/ Skull Pin 2-1/4in TI Sterl 1PK; 516V2S/ Skull Pin 3in TI Sterl 1PK; 520TS/ Skull Pin 2-1/4in TI Sterl 5PK; 520V2S/ Skull Pin 3in TI Sterl 5 PK

Reason

Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.

Code Info

All lots of the products manufactured from 06/30/2009 - 12/13/2018 Titanium Pin Label Mistakenly lists Ceramic Tip, Part/Lot numbers: 516TS/ AL18021418 and AL17022109; 520TS/ AL18021417 and AL15091012

Distribution

U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium

Quantity

372