FDA Enforcement Class II Terminated

Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

Recall: Z-0994-2015 · Reported January 28, 2015

Enforcement

Recall Number
Z-0994-2015
Event ID
70020
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 28, 2015
Initiation Date
December 15, 2014
Classification Date
January 16, 2015
Termination Date
January 26, 2017
Address
51 Valley Stream Pkwy, N/A, Malvern, PA, 19355-1406, United States

Description

Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

Reason

Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.

Code Info

Material 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 10762470 serial numbers24077 24051 24038 24059 24066 24015 24026 24058 24039 24090 24075

Distribution

US Distribution to states of: ID, IL, MI, MN, MO, ND and PA.

Quantity