FDA Enforcement Class II Terminated

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

Recall: Z-0993-2014 · Reported February 26, 2014

Enforcement

Recall Number
Z-0993-2014
Event ID
67076
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
The Standing Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 26, 2014
Initiation Date
May 16, 2013
Classification Date
February 19, 2014
Termination Date
June 25, 2014
Address
5848 Dixie Hwy, Saginaw, MI, 48601-5967, United States

Description

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

Reason

Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.

Code Info

Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427.

Distribution

Nationwide Distribution including AL, CA, DE, IA, IL , IN, KS, KY, MI, MO, NJ, OH, OK, TX, WA, WI,and WV.

Quantity

approximately 119