FDA Enforcement Class II Ongoing

Flow-i C40 Anesthesia System, model 6677400

Recall: Z-0989-2024 · Reported February 14, 2024

Enforcement

Recall Number
Z-0989-2024
Event ID
93736
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Getinge Usa Sales Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 14, 2024
Initiation Date
December 11, 2023
Classification Date
February 6, 2024
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035, United States

Description

Flow-i C40 Anesthesia System, model 6677400

Reason

Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.

Code Info

UDI-DI: 07325710001318, serial numbers (US): 5204, 5205, 5206, 5207, 5208, 5209, 5210, 5211, 7185, 7982, 7983, 8426, 8427, 8428, 8445, 8446.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.

Quantity

16 (US)