FDA Enforcement Class II Ongoing

Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q

Recall: Z-0989-2022 · Reported April 27, 2022

Enforcement

Recall Number
Z-0989-2022
Event ID
89825
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
St. Jude Medical, Cardiac Rhythm Management Division
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 27, 2022
Initiation Date
March 10, 2022
Classification Date
April 21, 2022
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577, United States

Description

Gallant VR Implantable Cardioverter Defibrillator REF # CDVRA500Q

Reason

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Code Info

REF# CDVRA500Q SN: 111009251, Exp Date : 31-Jul-2022, Material # 100173801; SN # 810000544, Exp Date: 31-Oct-2022, Material # 600115400 UDI# 05415067031990

Distribution

Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.

Quantity

2