FDA Enforcement Class II Terminated

Maquet XS Flat Screen Monitor Holder, PWD70+DF HD R XS32 K3 SPE, Configuration Part Number ARD268400150A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Recall: Z-0989-2019 · Reported March 13, 2019

Enforcement

Recall Number
Z-0989-2019
Event ID
82074
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GETINGE US SALES LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 13, 2019
Initiation Date
January 8, 2019
Classification Date
March 7, 2019
Termination Date
June 3, 2020
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

Maquet XS Flat Screen Monitor Holder, PWD70+DF HD R XS32 K3 SPE, Configuration Part Number ARD268400150A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Reason

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Code Info

Serial Numbers 500001 500002 500003 500004 500005 500006

Distribution

The products were distributed to the following US states: AZ and TX.

Quantity

6