FDA Enforcement Class II Ongoing

CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

Recall: Z-0988-2025 · Reported January 29, 2025

Enforcement

Recall Number
Z-0988-2025
Event ID
95602
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
St. Jude Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2025
Initiation Date
October 5, 2024
Classification Date
January 17, 2025
Address
387 Technology Cir Nw Ste 500, Atlanta, GA, 30313-2424, United States

Description

CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

Reason

As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.

Code Info

MN6000 (Which interfaces with CM1100 for PES) UDI-DI Code: 05414734510035 Merlin ID / PES Serial Number: 027556711 CM1100 P180912371 444058655 CM1100 P240100658 494955560 CM1100 P230300384 497887091 CM1100 P180208671 513914390 CM1100 P230800825 742148364 CM1100 P220800695 751954397 CM1100 P211200289 000697912 CM1100 P181013064 001545862 CM1100 P220400808,P220100189 003473584 CM1100 P211100353 005842000 CM1100 P230100314 006333918 CM1100 P220600729 008418758 CM1100 P240300492 010473700 CM1100 P210700575 010485845 CM1100 P220100057 011307774 CM1100 P190800198 013397427 CM1100 P220500363 018103333 CM1100 P210201417 018852830 CM1100 P190900022 019716455 CM1100 P221100752 020956348 CM1100 P230500703 021148283 CM1100 P160900736 021410246 CM1100 P230800171 023874951 CM1100 P230400238 024386088 CM1100 P200100300 026473926 CM1100 P240500531 031880776 CM1100 P240300135 032455370 CM1100 P240300138 034701771 CM1100 P230300580 037825221 CM1100 P240100040 039458344 CM1100 P190700191 042069666 CM1100 P221200307 043804384 CM1100 P220401019 044375798 CM1100 P200100451 046139531 CM1100 P210400699 046659165 CM1100 P220700471 049107915 CM1100 P220600696 049584410 CM1100 P230800848 053752093 CM1100 P230900101 053919908 CM1100 P200600006 054799366 CM1100 P211100584 055804058 CM1100 P180912650 056345895 CM1100 P210200735 057332710 CM1100 P240100823 058913997 CM1100 P240100775 060968161 CM1100 P240200495 061420626 CM1100 P230500071 062897822 CM1100 P240300373 066957044 CM1100 P210700838 067353847 CM1100 P240300708 067723346 CM1100 P240400015 068715747 CM1100 P220600972 069131084 CM1100 P200100430 069742013 CM1100 P230900257 069865251 CM1100 P230100100 073642522 CM1100 P230300687 074384090 CM1100 P200300463 076330232 CM1100 P210600422 077130896 CM1100 P230700227 077880912 CM1100 P240100412 079272753 CM1100 P240300622 081971756 CM1100 P221100962 082613308 CM1100 P230100076 083258111 CM1100 P230200980 083964213 CM1100 P240600232 087932182 CM1100 P200300339 088913132 CM1100 P210600328 089528947 CM1100 P231100056 091435883 CM1100 P230600230 091846907 CM1100 P190400092 092964808 CM1100 P220601351 093046498 CM1100 P180208408 096655907 CM1100 P200500395 097782585 CM1100 P220800056 097973184 CM1100 P210400470 097985105 CM1100 P220200308 098510456 CM1100 P180912257 098553217 CM1100 P230700024 098717770 CM1100 P180812191 103524443 CM1100 P190600308 105005185 CM1100 P240400232 106316250 CM1100 P240100164 110571163 CM1100 P231100263 110786878 CM1100 P200200279 111636379 CM1100 P191200030 112437421 CM1100 P190214131 112450366 CM1100 P220100031 114078702 CM1100 P220400840 114746951 CM1100 P220400246 115415655 CM1100 P200600103 117402925 CM1100 P200200448 121663629 CM1100 P240400540 123570731 CM1100 P240300559 124446626 CM1100 P220600722 125968610 CM1100 P220300255 129028643 CM1100 P240500155 129187100 CM1100 P240100092 130416936 CM1100 P220300540 130561434 CM1100 P220500091 131755126 CM1100 P240600068 134652791 CM1100 P211200310 135234896 CM1100 P180811660 135686780 CM1100 P220400411 136869880 CM1100 P231000033 139677223 CM1100 P190700508 140298902 CM1100 P210700639 140798877 CM1100 P201000097 142033547 CM1100 P210100076 146549621 CM1100 P240300424 146795588 CM1100 P230500077 154283196 CM1100 P210300069 158101204 CM1100 P230500393 162001218 CM1100 P190700311 165130402 CM1100 P240600002 168144863 CM1100 P240300776 168168185 CM1100 P240100786 168785889 CM1100 P220700465 173179722 CM1100 P231100242 178994091 CM1100 P200200206 181681214 CM1100 P230200164 182135285 CM1100 P220300192 182232033 CM1100 P230300422 183267517 CM1100 P220400387 184558245 CM1100 P220600457 184559425 CM1100 P240100020 186511887 CM1100 P230900092 190528869 CM1100 P180610764 190891739 CM1100 P230400467 193168929 CM1100 P180108014 194501615 CM1100 P220300300 194551065 CM1100 P210100404 195809595 CM1100 P170604562 195966452 CM1100 P220700188 197184963 CM1100 P221100939 198990079 CM1100 P180812065 199200189 CM1100 P240100007 200395739 CM1100 P200200308 202836730 CM1100 P240100179 208758755 CM1100 P240400516 208772129 CM1100 P190214217 212681613 CM1100 P220600229 213719610 CM1100 P230700075 214403909 CM1100 P200200410 219149465 CM1100 P220200020 220570311 CM1100 P230600741 220801765 CM1100 P240300196 221186828 CM1100 P211000415 222745481 CM1100 P180309231 223692831 CM1100 P180409883 225728211 CM1100 P181012711 227541125 CM1100 P180610591 228372231 CM1100 P161000975 230252645 CM1100 P240100056 230334013 CM1100 P210400078 232604975 CM1100 P230600189 233939131 CM1100 P220600855 236466009 CM1100 P200600101 236768586 CM1100 P171006650 237203971 CM1100 P230800716 237263793 CM1100 P240400420 237987359 CM1100 P240900138 244016986 CM1100 P220500693 245197025 CM1100 P180711129 246163364 CM1100 P220500054 246340731 CM1100 P210700889 252307665 CM1100 P190113705 252386115 CM1100 P240500320 253106405 CM1100 P230200684 253613996 CM1100 P231000068 253830137 CM1100 P231000140 254778707 CM1100 P240200507 255438160 CM1100 P240100203 256944752 CM1100 P191000257 258523372 CM1100 P220601079 261777874 CM1100 P190800428 262395304 CM1100 P240300371 262778772 CM1100 P220200261 263605636 CM1100 P210900193 266524180 CM1100 P210500010 268267234 CM1100 P230200053 268484839 CM1100 P200600222 270074206 CM1100 P171106968 274979970 CM1100 P221100675 275289833 CM1100 P200100099 277187068 CM1100 P210600057 277220257 CM1100 P220200431 279610018 CM1100 P221100001 280474222 CM1100 P180812087 283021137 CM1100 P221100156 283696672 CM1100 P200500400 284467206 CM1100 P240100344 284692092 CM1100 P210700044 292458718 CM1100 P220700181 294186242 CM1100 P180710945 296152457 CM1100 P180912276 298042227 CM1100 P200500172 298070806 CM1100 P220800495 300912730 CM1100 P231000261 305500472 CM1100 P220500805 305819500 CM1100 P240200437 306426487 CM1100 P230800796 306758319 CM1100 P210700851 307202176 CM1100 P220300431 307880484 CM1100 P210600155 309781797 CM1100 P211200233 310695366 CM1100 P211000748 311217103 CM1100 P220601485 311631733 CM1100 P191000242 311934103 CM1100 P240300491 319975595 CM1100 P230500579 320029176 CM1100 P180308923 320602477 CM1100 P200200215 322892183 CM1100 P230400723 329333967 CM1100 P220800470 331918607 CM1100 P240300444 343355574 CM1100 P191100372 344087887 CM1100 P211200547 344588140 CM1100 P190113708 346926272 CM1100 P230400491 347608044 CM1100 P191000064 348104514 CM1100 P211000874 352162697 CM1100 P240800667 352939748 CM1100 P221100887 354086183 CM1100 P240300177 354240673 CM1100 P161001101 356395038 CM1100 P170202830 356952820 CM1100 P240200190 359439601 CM1100 P240400177 359865490 CM1100 P210700105 364505818 CM1100 P171207516 368265583 CM1100 P230400431 369124821 CM1100 P230400020 369157375 CM1100 P230800001 370123093 CM1100 P200200187 370274961 CM1100 P220300326 371594318 CM1100 P210900057 372379966 CM1100 P180309127 373440346 CM1100 P230800140 375911823 CM1100 P230800741 376518650 CM1100 P240200198 377185616 CM1100 P220400244 382727717 CM1100 P180409696 385693189 CM1100 P210900037 388377350 CM1100 P181113383 389104423 CM1100 P230800757 391036423 CM1100 P230400497 395640220 CM1100 P180510254 401490206 CM1100 P230900171 402109201 CM1100 P190214182 402255517 CM1100 P170604828 402628804 CM1100 P200100587 403698244 CM1100 P201000162 404279978 CM1100 P230500069 406139915 CM1100 P240500400 406381392 CM1100 P180208619 408445914 CM1100 P170303208 408786705 CM1100 P220400372 409199106 CM1100 P240600394 411912546 CM1100 P210600320 413136805 CM1100 P220600868 413733015 CM1100 P221100197 414185025 CM1100 P240300040 416763076 CM1100 P240500097 419042015 CM1100 P230500146 423411206 CM1100 P240300333 426439923 CM1100 P230300045 426469698 CM1100 P181213482 426530952 CM1100 P240300417 427359195 CM1100 P161001007 428398739 CM1100 P211001056 429436892 CM1100 P230500337 431784867 CM1100 P240200236 432437499 CM1100 P220200276 435690169 CM1100 P190500328 436517833 CM1100 P190800471 437546872 CM1100 P231000304 439719758 CM1100 P171006206 441354826 CM1100 P240500341 442096970 CM1100 P190600120 444494447 CM1100 P230200181 447222555 CM1100 P210400103 449437243 CM1100 P230800031 451748636 CM1100 P240300335 453264376 CM1100 P210201464 458406519 CM1100 P240300775 461591992 CM1100 P230900435 462620105 CM1100 P180308803 464625540 CM1100 P230200657 467654000 CM1100 P230100489 470268780 CM1100 P240200471 478579014 CM1100 P230300065 479067118 CM1100 P220100317 480351741 CM1100 P240300777 482431962 CM1100 P190400060 482786886 CM1100 P191000323 483625521 CM1100 P210400623 485241988 CM1100 P220800322 486068760 CM1100 P211000369 489189498 CM1100 P220500635 489545228 CM1100 P200300547 490915188 CM1100 P161101333 491185856 CM1100 P191000488 492192455 CM1100 P180308850 494410608 CM1100 P230300210 497905380 CM1100 P171006384 500487129 CM1100 P210700039 501371454 CM1100 P171207485 502428865 CM1100 P240100213 502652464 CM1100 P240300751 503103988 CM1100 P240100035 505457945 CM1100 P220400441 505812966 CM1100 P220400326 508809456 CM1100 P240200163 514258540 CM1100 P221100638 514742907 CM1100 P191100429 515592285 CM1100 P190400108 518863527 CM1100 P240300379 518872569 CM1100 P190214238 520502048 CM1100 P190600463 523073526 CM1100 P210500194 523413813 CM1100 P230100030 525600276 CM1100 P230100618 528575541 CM1100 P231100039 529075954 CM1100 P220601113 531902971 CM1100 P220100442 533161279 CM1100 P210200287 537268872 CM1100 P210100241 541387478 CM1100 P201200110 542890504 CM1100 P230600172 544878077 CM1100 P240500279 548079920 CM1100 P180912565 548337351 CM1100 P220400170 554641290 CM1100 P230200905 559011242 CM1100 P220100084 563392091 CM1100 P200100185 563393867 CM1100 P240400032 565899812 CM1100 P240300749 566845483 CM1100 P220800221 567619564 CM1100 P180912421 570331462 CM1100 P240400315 574608816 CM1100 P171006324 575367487 CM1100 P190500320 575751227 CM1100 P190500218 576525729 CM1100 P190400314 576755615 CM1100 P171006229 579037078 CM1100 P240300013 579917048 CM1100 P191100200 581159357 CM1100 P240400434 584384242 CM1100 P200800193 585309040 CM1100 P240100144 585385362 CM1100 P240100500 585776347 CM1100 P210500150 587345547 CM1100 P181113137 588282160 CM1100 P210300707 591516646 CM1100 P220100583 591582986 CM1100 P180409773 592241806 CM1100 P220601266 592748172 CM1100 P220800352 593964315 CM1100 P231000284 595785213 CM1100 P190113867 598081636 CM1100 P230800062 598880136 CM1100 P230300544 599754991 CM1100 P240200125 604960369 CM1100 P200100389 605049881 CM1100 P210100550 607262144 CM1100 P200500212 608065025 CM1100 P240300612 608360426 CM1100 P170604566 609922372 CM1100 P240800612 611670993 CM1100 P230500624 612076232 CM1100 P200100440 614992014 CM1100 P240600470 616754834 CM1100 P221100149 617507090 CM1100 P240300691 619439599 CM1100 P240100460 621482165 CM1100 P240200003 622754570 CM1100 P221100870 624194502 CM1100 P240100148 624680724 CM1100 P220800734 625199443 CM1100 P191000264 628310179 CM1100 P190900429 629565268 CM1100 P230700303 631102530 CM1100 P230500407 631744810 CM1100 P191000173 632858882 CM1100 P230300575 635604564 CM1100 P230100090 635758147 CM1100 P190500437 636872905 CM1100 P170202924 639572148 CM1100 P230600726 642304745 CM1100 P230900155 645873274 CM1100 P221000242 648482446 CM1100 P201200026 651159881 CM1100 P200600159 655100725 CM1100 P200500359 655194108 CM1100 P221100810 655610533 CM1100 P211000805 658356548 CM1100 P230200048 661284158 CM1100 P210800277 662122597 CM1100 P230300434 664211844 CM1100 P230500364 664481181 CM1100 P240100563 665295333 CM1100 P170403567 665876389 CM1100 P240400554 665951562 CM1100 P181012699 666566716 CM1100 P210201742 667673305 CM1100 P211000032 670348416 CM1100 P230300121 671566222 CM1100 P200100645 671689248 CM1100 P220700363 673361846 CM1100 P230200471 674022702 CM1100 P231000282 674381496 CM1100 P240400424 675903546 CM1100 P230300345 677916777 CM1100 P220700442 678702556 CM1100 P171006174 680165248 CM1100 P240600437 681661054 CM1100 P240200022 683291496 CM1100 P230800016 684586274 CM1100 P240300405 685343600 CM1100 P230800158 686032103 CM1100 P230600743 686933599 CM1100 P220600726 689708394 CM1100 P180711232 694165630 CM1100 P240300107 698752292 CM1100 P230200152 699163853 CM1100 P211000784 705151157 CM1100 P211100260 708002563 CM1100 P220600996 714184215 CM1100 P200100698 714517281 CM1100 P210300149 714829553 CM1100 P230500546 715261301 CM1100 P211001008 718077282 CM1100 P180711086 718418049 CM1100 P171207239 719691503 CM1100 P210100193 722379310 CM1100 P200200361 723386686 CM1100 P180510142 724180369 CM1100 P240300240 724925474 CM1100 P180610748 726376833 CM1100 P220800542 726938780 CM1100 P230300563 728463399 CM1100 P240100033 728530320 CM1100 P220200171 731136727 CM1100 P190500548 732306675 CM1100 P230400237 734375009 CM1100 P210201559 735883530 CM1100 P200900387 737254938 CM1100 P210400320 737796664 CM1100 P180409637 740953146 CM1100 P221100734 740962196 CM1100 P221100457 741020838 CM1100 P230600070 744500802 CM1100 P230900100 746102680 CM1100 P221100409 746304625 CM1100 P230500581 746925627 CM1100 P230200394 747548667 CM1100 P170303060 749321931 CM1100 P220300815 749821815 CM1100 P230200914 752835280 CM1100 P210500659 753829977 CM1100 P240700553 754398147 CM1100 P220400575 754401651 CM1100 P181213603 755657681 CM1100 P230800720 759306988 CM1100 P210400182 759483738 CM1100 P220600324 760200477 CM1100 P230400051 762019446 CM1100 P220500573 765348461 CM1100 P211100040 766713606 CM1100 P190500513 767078728 CM1100 P240100167 768139438 CM1100 P191000050 770350551 CM1100 P240300004 774312334 CM1100 P220300393 776738049 CM1100 P220601421 778414466 CM1100 P200300178 779034875 CM1100 P211000430 783494818 CM1100 P180107859 784325144 CM1100 P231000146 788828580 CM1100 P230800060 789854882 CM1100 P210400421 792437352 CM1100 P220100233 792902140 CM1100 P161201586 794009647 CM1100 P211000469 794891044 CM1100 P200100677 794927616 CM1100 P171207240 795221381 CM1100 P211100169 795353622 CM1100 P200900633 796726867 CM1100 P240600616 797146859 CM1100 P210300001 797869492 CM1100 P230400007 798774089 CM1100 P220600528 799114012 CM1100 P240300165 800172652 CM1100 P230900212 801440314 CM1100 P220700435 802016253 CM1100 P240200213 806589164 CM1100 P221100025 807772652 CM1100 P200500094 808438337 CM1100 P240400205 810120527 CM1100 P220200566 811163039 CM1100 P210900191 811515519 CM1100 P240100794 814538922 CM1100 P240200485 818718033 CM1100 P231000060 819628181 CM1100 P240500374 819846346 CM1100 P240300394 821261351 CM1100 P200300202 821395183 CM1100 P210100061 823485537 CM1100 P220300653 823780572 CM1100 P230300419 824955009 CM1100 P210500361 826531626 CM1100 P171006348 832456420 CM1100 P220100526 833440134 CM1100 P190700052 835022484 CM1100 P240700269 838296556 CM1100 P170504373 838558856 CM1100 P191000492 839509684 CM1100 P170403585 842318214 CM1100 P200900402 843784299 CM1100 P190400047 845166248 CM1100 P190700289 846802379 CM1100 P240400160 848083861 CM1100 P240100080 848897336 CM1100 P180409536 849440722 CM1100 P220100549 849732680 CM1100 P161101315 851163501 CM1100 P240300456 851958959 CM1100 P230700304 853122125 CM1100 P210600265 854198652 CM1100 P180711361 855136610 CM1100 P200200186 858261464 CM1100 P240100524 859356164 CM1100 P230900173 860594332 CM1100 P240100102 860636729 CM1100 P180308876 860806231 CM1100 P171006506 864374137 CM1100 P230200347 866807167 CM1100 P210500093 867567091 CM1100 P200100198 871612974 CM1100 P211200619 872071428 CM1100 P220200507 875710329 CM1100 P221100302 878038850 CM1100 P200200179 878705516 CM1100 P211100385 883067779 CM1100 P220500205 883727505 CM1100 P240400561 890828999 CM1100 P210701016 890843410 CM1100 P200900191 893265843 CM1100 P231000213 902771120 CM1100 P200300207 909873747 CM1100 P240200429 910094788 CM1100 P200900608 911005163 CM1100 P170403513 911127033 CM1100 P200200065 911591865 CM1100 P220500819 914260559 CM1100 P240100239 914553300 CM1100 P230300447 917161101 CM1100 P240100733 919430611 CM1100 P220600906 920128998 CM1100 P231100270 922495528 CM1100 P220200029 924328610 CM1100 P230400078 925751638 CM1100 P220600187 925927089 CM1100 P190600364 926934803 CM1100 P240300651 927265959 CM1100 P191100060 930980545 CM1100 P240400458 931322200 CM1100 P220400922 936060623 CM1100 P240200096 938609021 CM1100 P230700237 939617007 CM1100 P170504272 940480338 CM1100 P240300433 940521495 CM1100 P170303264 940932684 CM1100 P170202310 941742918 CM1100 P211000087 942358482 CM1100 P190700244 943421438 CM1100 P230800828 946803467 CM1100 P221100125 948067095 CM1100 P220300245 949476246 CM1100 P230400673 950677823 CM1100 P240300790 950772632 CM1100 P230200723 953281755 CM1100 P240200067 953715968 CM1100 P210500191 955545645 CM1100 P220400831 962725750 CM1100 P221000207 962980850 CM1100 P240300476 965949571 CM1100 P230500704 971509831 CM1100 P230400215 974042541 CM1100 P230700057 976587576 CM1100 P170805497 977722453 CM1100 P220200203 978030971 CM1100 P230800110 978664092 CM1100 P211200166 978757540 CM1100 P190700603 979688181 CM1100 P190214450 979999620 CM1100 P221100659 980222624 CM1100 P230800136 981499353 CM1100 P220900133 981833593 CM1100 P180208707 986810596 CM1100 P191000301 988054110 CM1100 P240400522 988866935 CM1100 P210700912 992239657 CM1100 P230200873 995330511 CM1100 P230800017

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NV, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and the countries of Australia, and Canada.

Quantity

605 units