FDA Enforcement Class II Terminated

Cocoon Convective Warming System, Product Code: CWS5000 (110V).

Recall: Z-0987-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-0987-2020
Event ID
84541
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Care Essentials Pty., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
February 12, 2020
Initiation Date
January 7, 2020
Classification Date
February 4, 2020
Termination Date
March 22, 2023
Address
25 Slevin St, N/A, North Geelong, N/A, N/A, Australia

Description

Cocoon Convective Warming System, Product Code: CWS5000 (110V).

Reason

Potential for thermal damage near the power cord connector and Power Entry Module (PEM) due to improper connection of the power cord connector and PEM pins on the Cocoon Convective Warming System.

Code Info

Product Code: CWS5000 (110V) Serial Numbers: CWS5-400001-18, CWS5-400011-18, S00001, S00003, S00004, S00006, S00007, S00008, S00009, S00010, S00011, S00012, S00013, S00014, S00015, S00016, S00017, S00018, S00019, S00020, S00021, S00022, S00023, S00024, S00025, S00026, S00027, S00028, S00029, S00030, S00031, S00032, S00033, S00034, S00035, S00036, S00037, S00038, S00039, S00040, S00041, S00042, S00043, S00044, S00045, S00046, S00047, S00048, S00049, S00050, S00051, S00052, S00053, S00054, S00055, S00056, S00057, S00058, S00059, S00060, S00061, S00062, S00063, S00064, S00065, S00066, S00067, S00068, S00069, S00070, S00071, S00072, S00073, S00074, S00075, S00076, S00077, S00078, S00079, S00080, S00081, S00082, S00083, S00084, S00085, S00086, S00087, S00088, S00089, S00090, S00091, S00092, S00093, S00094, S00095, S00096, S00098, S00099, S00100, S00101, S00102, S00103, S00104, S00105, S00106, S00107, S00108, S00109, S00110, S00111, S00112, S00113, S00114, S00115. S00116, S00117, S00118, S00119, S00120, S00121, S00122, S00123, S00124, S00125, S00126, S00127, S00128, S00129, S00130, S00131, S00132, S00133, S00134, S00135, S00136, S00137, S00138, S00139, S00140, S00141, S00142, S00143, S00144, S00145, S00146, S00147, S00148, S00149, S00150, S00151, S00152, S00153, S00154, S00155, S00156, S00157, S00158, S00159, S00160, S00161, S00162, S00163, S00164, S00165, S00166, S00167, S00168, S00169, S00170, S00171, S00172, S00173, S00174, S00175, S00176, S00177, S00178, S00179, S00180, S00181, S00182, S00183, S00184, S00185, S00186, S00187, S00188, S00189, S00190, S00191, S00192, S00193, S00194, S00195, S00196, S00197, S00198, S00199, S00200, S00201, S00202, S00203, S00204, S00205, S00206, S00207, S00208, S00209, S00210, S00211, S00212, S00213, S00214, S00215, S00216, S00217, S00218, S00219, S00220, S00221, S00222, S00223, S00224, S00225, S00226, S00227, S00228, S00229, S00230, S00231, S00232, S00233, S00234, S00235, S00236, S00237, S00238, S00239, S00240, S00241, S00242, S00243, S00244, S00245, S00246, S00247, S00248, S00249, S00250, S00251, S00252

Distribution

US Nationwide in the states of AZ CA FL GA KS LA MA MD MI MN NC NM NY OH PA QC SC TX VA WV and the Foreign distribution to Canada.

Quantity

251