FDA Enforcement
Class II
Terminated
Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Recall: Z-0987-2019
·
Reported March 13, 2019
Enforcement
- Recall Number
- Z-0987-2019
- Event ID
- 82074
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GETINGE US SALES LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 13, 2019
- Initiation Date
- January 8, 2019
- Classification Date
- March 7, 2019
- Termination Date
- June 3, 2020
- Address
- 45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States
Description
Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Reason
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Code Info
Serial Numbers 500009 500010 500011 500012 500013 500014
Distribution
The products were distributed to the following US states: AZ and TX.
Quantity
6