FDA Enforcement Class II Ongoing

DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,

Recall: Z-0982-2024 · Reported February 14, 2024

Enforcement

Recall Number
Z-0982-2024
Event ID
93724
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 14, 2024
Initiation Date
November 21, 2023
Classification Date
February 5, 2024
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037, United States

Description

DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,

Reason

There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2) System provides numerical results that are below the lowest reportable result if laboratory chooses to report assay results in units of measurement other than the defined default units; 3) A result of 0 incorrectly reported when the system utilizes an expression that utilizes non-numerical symbols (e.g., > or <) to derive a calculated test as part of result reporting; 4) Reserve volume is enabled and the assay LIS code is not the same as the assay Test ID. These issues could potentially lead to erroneous results or delay reporting results for multiple analytes.

Code Info

UDI/DI 15099590732103, Serial Numbers: 300116, 300117, 300118, 300123, 300127, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300138, 300139, 300137, 300140, 300142, 300141, 300144, 300143, 300145, 300148 300146, 300147, 300149, 300150, 300151, 300152, 300155, 300153, 300154, 300156, 300158, 300160, 300163, 300161, 300162, 300164, 300167, 300170, 300169, 300171, 300168, 300166, 300172, 300176, 300173, 300178, 300181, 300179, 300183, 300182, 300188, 300189, 300190, 300191, 300192, 300195, 300196, 300193, 300194, 300197, 300199, 300198, 300201, 300200, 300202, 300203, 300205, 300204 300207, 300206, 300208, 300209, 300212, 300210, 300211, 300213, 300214, 300215, 300217, 300216, 300220, 300218, 300219, 300221, 300223, 300224, 300222, 300227, 300228, 300229, 300231, 300230, 300235, 300236, 300234, 300237, 300232, 300239, 300233, 300238, 300240, 300243, 300241, 300242, 300247, 300246, 300249, 300253, 300244, 300250, 300252, 300248, 300251, 300256, 300257, 300254, 300258, 300255, 300259, 300260, 300261, 300264, 300263, 300262; running software version 1.16 or prior

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Croatia, Czech Republic, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

Quantity

128 analyzers