FDA Enforcement
Class II
Terminated
Anchored Lateral Retractable Drill, Timberline IPF IFU IFU8621-0113, Model 8630-0102. Timberline Anchored Lateral Retractable Drills are used with the Timberline MPF System to establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.
Recall: Z-0982-2015
·
Reported January 21, 2015
Enforcement
- Recall Number
- Z-0982-2015
- Event ID
- 70056
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet Spine, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 21, 2015
- Initiation Date
- August 22, 2013
- Classification Date
- January 15, 2015
- Termination Date
- March 13, 2015
- Address
- 310 Interlocken Pkwy Ste 120, Broomfield, CO, 80021-3464, United States
Description
Anchored Lateral Retractable Drill, Timberline IPF IFU IFU8621-0113, Model 8630-0102. Timberline Anchored Lateral Retractable Drills are used with the Timberline MPF System to establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.
Reason
The outer diameter of the Drill shaft is oversized; thereby, resulting in interference fit when inserting the Drill into the Fixed or Variable Sleeve Assembly.
Code Info
Model 8630-0102 Anchored Lateral Retractable Drill lot TU00049: February 25, 2013 and TU00155: July 31, 2013
Distribution
Distributed in PR and the states of NY, AZ, TN, TX, and CA.
Quantity
15