FDA Enforcement Class II Terminated

BD MAX Enteric Bacterial Panel -For In Vitro Diagnostic Use with the BD MAX System Catalog Number: 442963 The BD MAX Enteric Bacterial Panel performed on the BD MAX System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens.

Recall: Z-0981-2019 · Reported March 13, 2019

Enforcement

Recall Number
Z-0981-2019
Event ID
82046
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 13, 2019
Initiation Date
January 14, 2019
Classification Date
March 6, 2019
Termination Date
April 3, 2020
Address
7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States

Description

BD MAX Enteric Bacterial Panel -For In Vitro Diagnostic Use with the BD MAX System Catalog Number: 442963 The BD MAX Enteric Bacterial Panel performed on the BD MAX System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric bacterial pathogens.

Reason

False negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides targets of the BD MAX Extended Enteric Bacterial Panel assay.

Code Info

Lot number/UDI 8194784 (01)00382904429638 (17)200222(10)8194784(30)1 and 8199682 (01)00382904429638 (17)200222(10)8199682(30)1 Expiration Date: 2020-02-22

Distribution

Worldwide Distribution: US (nationwide) to states of:CA, GA, HI, ID, MI, MN, MT, NJ, NM, OH, OR, PA, SC, TX, VA, VT, WA ; and to countries of: Australia, New Zealand, Canada, Europe, and Korea.

Quantity

1583 assay kits