FDA Enforcement Class II Ongoing

Pacific Medical GE Corometrics Nautilus Toco Transducer w/wing (2264LAX), FFCM3260

Recall: Z-0980-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-0980-2020
Event ID
83869
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Pacific Medical Group Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2020
Initiation Date
September 19, 2019
Classification Date
February 4, 2020
Address
212 Avenida Fabricante, N/A, San Clemente, CA, 92672-7538, United States

Description

Pacific Medical GE Corometrics Nautilus Toco Transducer w/wing (2264LAX), FFCM3260

Reason

Fetal transducers distributed in advance of receiving 510(k) and international clearances. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and interruption of monitoring due to device failure that could lead to delay in detecting maternal or fetal distress.

Code Info

All serial numbers distributed between 02/06/16 - 04/30/19.

Distribution

U.S.: IN, AZ, IL, HI, AK, AR, CA, FL, GA, IA, ID, LA, MA, NC, NE, NM, NY, OK, PA, TN, TX, UT, VA, WA, WV, NV, KY, CO, ME, ND, OH, SC, WI, AL, OR, DC, DE, MD, MI, MN, MO, MT, MS, NJ, NH. O.U.S. (Foreign): CA, CO, FR, GB, IT, MX, UAE

Quantity

628