FDA Enforcement
Class I
Terminated
Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364. For transseptal left heart access in both diagnostic and interventional procedures.
Recall: Z-0979-2019
·
Reported March 20, 2019
Enforcement
- Recall Number
- Z-0979-2019
- Event ID
- 82124
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 20, 2019
- Initiation Date
- February 1, 2019
- Classification Date
- March 13, 2019
- Termination Date
- June 30, 2020
- Address
- 750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States
Description
Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364. For transseptal left heart access in both diagnostic and interventional procedures.
Reason
The products were manufactured without a back bevel on the needle tip. This could cause damage to the inside of the introducer during needle insertion.
Code Info
Lot 8833687
Distribution
US nationwide and Canada. US distribution to the following states: AZ, FL, IL, IN, KS, MA, MI, MO, PA, SC, TN, and WA. The products were distributed to the following foreign countries: Canada.
Quantity
100