FDA Enforcement Class I Terminated

e700 Automatic Transport Ventilator, Model Number 01EVE700

Recall: Z-0978-2019 · Reported April 10, 2019

Enforcement

Recall Number
Z-0978-2019
Event ID
82178
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
O-Two Medical Technologies, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 10, 2019
Initiation Date
February 12, 2019
Classification Date
April 4, 2019
Termination Date
June 29, 2023
Address
45A Armthorpe Road, N/A, Brampton, Ontario, N/A, Canada

Description

e700 Automatic Transport Ventilator, Model Number 01EVE700

Reason

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Code Info

EV70001 to EV70658

Distribution

The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.

Quantity

632