FDA Enforcement
Class II
Terminated
Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 86300204 Timberline Anchored Lateral Variable Retractable Sleeves (the Sleeves) are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.
Recall: Z-0978-2015
·
Reported January 21, 2015
Enforcement
- Recall Number
- Z-0978-2015
- Event ID
- 70060
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet Spine, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 21, 2015
- Initiation Date
- August 22, 2013
- Classification Date
- January 14, 2015
- Termination Date
- March 13, 2015
- Address
- 310 Interlocken Pkwy Ste 120, Broomfield, CO, 80021-3464, United States
Description
Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 86300204 Timberline Anchored Lateral Variable Retractable Sleeves (the Sleeves) are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.
Reason
The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.
Code Info
Model 8630-0204 Lots TU00052, TU00052A, and TU00117
Distribution
US Distribution to states of: AZ, CA, NY, TN, and TX including PR.
Quantity
12 units