FDA Enforcement Class II Ongoing

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609360 Software Version: N/A Product Description: Affixus 9 mm, Length 360 mm, Left, Trochanteric Nail Component: N/A

Recall: Z-0977-2026 · Reported December 31, 2025

Enforcement

Recall Number
Z-0977-2026
Event ID
98069
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Zimmer, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 31, 2025
Initiation Date
December 2, 2025
Classification Date
December 22, 2025
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609360 Software Version: N/A Product Description: Affixus 9 mm, Length 360 mm, Left, Trochanteric Nail Component: N/A

Reason

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Code Info

Lot Code: Model No 815609360 ; UDI-DI (01)00887868144493(17)340820(10)66717603 ; Lot Number 66717603 Model No 815609360 ; UDI-DI (01)00887868144493(17)340905(10)66892902 ; Lot Number 66892902 Model No 815609360 ; UDI-DI (01)00887868144493(17)341030(10)67031550 ; Lot Number 67031550

Distribution

US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Quantity

24