FDA Enforcement Class II Ongoing

Artis Q zeego, Model Number 10848283

Recall: Z-0977-2022 · Reported April 27, 2022

Enforcement

Recall Number
Z-0977-2022
Event ID
89860
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 27, 2022
Initiation Date
February 23, 2022
Classification Date
April 21, 2022
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

Artis Q zeego, Model Number 10848283

Reason

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Code Info

UDI: 4056869010007; Serial Numbers: 117021 117136 117324 117108 117001 117023 117329 117315 117125 117006 117312 117124 117129 117112 117034 117026 117029 117123 117100 117307 117117 117111 117304 117024 117325 117303 117030 117314 117309 117113 117322 117014 117040 117041 117114 117009 Additional Serial Numbers as of 10/4/23: 117127 117310

Distribution

US Nationwide.

Quantity

38 systems