e500 Automatic Transport Ventilator, Model Number 01EVE500
Recall: Z-0976-2019
·
Reported April 10, 2019
Enforcement
- Recall Number
- Z-0976-2019
- Event ID
- 82178
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- O-Two Medical Technologies, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 10, 2019
- Initiation Date
- February 12, 2019
- Classification Date
- April 4, 2019
- Termination Date
- June 29, 2023
- Address
- 45A Armthorpe Road, N/A, Brampton, Ontario, N/A, Canada
Description
e500 Automatic Transport Ventilator, Model Number 01EVE500
Reason
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Code Info
EV50001 to EV50148
Distribution
The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Quantity
136