FDA Enforcement Class II Ongoing

BD Pyxis Med ES Auxiliary (AUX), REF 343

Recall: Z-0975-2025 · Reported January 29, 2025

Enforcement

Recall Number
Z-0975-2025
Event ID
96053
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
CareFusion 303, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 29, 2025
Initiation Date
February 14, 2023
Classification Date
January 17, 2025
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386, United States

Description

BD Pyxis Med ES Auxiliary (AUX), REF 343

Reason

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Code Info

UDI/DI 10885403512698, Serial Numbers: 16171753, 16191650, 16192894, 16144435, 16154605, 16154606, 16154607, 16191713, 16191714, 16191715, 16191716, 16191718, 16191719, 16191720, 16191721, 16191722, 16196167, 16186595, 16196168, 16196169, 16186597, 16196170, 16196172, 16163310, 16179747, 16168959, 16167684, 16163677, 16185817, 16188363, 16188367, 16188695, 16185802, 16188322, 16191726, 16173525, 16173527, 16173528, 16194920, 16194922, 16190720, 16186580, 16191729, 16191730, 16191876, 16191877, 16191878, 16191879, 16166841, 16166842, 16182585, 16183929, 16178220, 16178222, 16179754, 16179750, 16179751, 16179752, 16179756, 16179757, 16179758, 16179759, 16179761, 16179762, 16184811, 16179764, 16179753, 16179765, 16179766, 16179767, 16179768, 16191682, 16191683, 16191684, 16188702, 16188705, 16172879, 16190710, 16190711, 16177195, 16177196, 16177199, 16177368, 16177369, 16191880, 16191881, 16188361, 16188364, 16188694, 16189315, 16173027, 16180723, 16173028, 16162674, 16186590, 16166843, 16177168, 16163837, 16172982, 16172983, 16164738, 16164740, 16172987, 16172989, 16166814, 16191646, 16182586, 16178208, 16179763, 16154507, 16154508, 16168957, 16147503, 16188768, 16191685, 16190713, 16197442, 16188355, 16196904, 16196905, 16196906, 16196907, 16196910, 16196916, 16192954, 16192955, 16192956, 16185809, 16192957, 16185810, 16185811, 16185812, 16188765, 16203246, 16126735, 16196903, 16196908, 16196915, 16196917, 16178206, 16178207, 16185544, 16180801, 16196932, 16184893, 16184894, 16188359, 16188360, 16188362, 16188365, 16188697, 16188735, 16172984, 16172986, 16172990, 16172991, 16172992, 16161338, 16197445, 16197446, 16197448, 16197449, 16163319, 16172880, 16154485, 16154486, 16154498, 16154499, 16154382, 16154383, 16154384, 16154466, 16154470, 16154504, 16154506, 16150226, 16150243, 16199333, 16199334, 16199335, 16199336, 16199337, 16199338, 16199339, 16199340, 16199341, 16199342, 16199343, 16199344, 16199345, 16199346, 16199347, 16199348, 16199349, 16199350, 16199351, 16199352, 16199353, 16199354, 16176443, 16176444, 16165142, 16165143, 16172909, 16172911, 16172934, 16203260, 16203261, 16175778, 16173007, 16192557, 16163794, 16166836, 16180711, 16190734, 16191697, 16191698, 16191699, 16191700, 16191701, 16185792, 16191706, 16191707, 16191708, 16191882, 16191709, 16140164, 16140171, 16186583, 16186585, 16163802, 16163803, 16163805, 16194923, 16194925, 16194926, 16194928, 16176458, 16196864, 16196866, 16173058, 16173060, 16175760, 16173002, 16173004, 16179975, 16173005, 16173006, 16173008, 16181359, 16168973, 16122586, 16197444, 16197447, 16198607, 16198608, 16198609, 16204117, 16163215, 16166837, 16188741, 16188746, 16154754, 16150261, 16149841, 16150287, 16147736, 16154648, 16050400, 16050401, 16182594, 16182595, 16182596, 16182598, 16182600, 16182601, 16182603, 16182616, 16183957, 16188700, 16189226, 16183908, 16185675, 16183909, 16183911, 16183912, 16183913, 16183914, 16149856, 16178210, 16178211, 16178212, 16178214, 16178217, 16178219, 16178221, 16178223, 16178224, 16178225, 16178226, 16176440, 16176441, 16176442, 16168974, 16168975, 16177164, 16133693, 16133694, 16186588, 16198595, 16198597, 16198598, 16198599, 16198600, 16198601, 16198602, 16198604, 16188356, 16188357, 16185602, 16199372, 16196152, 16196153, 16176367, 16178165, 16198625, 16198626, 16198628, 16198629, 16198630, 16198632, 16198633, 16198634, 16198635, 16198636, 16184896, 16188739, 16171754, 16135308, 16171765, 16135311, 16188740, 16188766, 16194924, 16194927, 16194929, 16194931, 16184178, 16188699, 16188736, 16188737, 16163205, 16163206, 16163207, 16163208, 16188742, 16188744, 16188767, 16185800, 16185801, 16191884, 16185803, 16191903, 16185805, 16196902, 16196909, 16196911, 16196912, 16205653, 16204132, 16204133, 16204135, 16204137, 16176378, 16205187, 16201121, 16201123, 16189294, 16189295, 16134551, 16177366, 16177367, 16177372, 16177373, 16177376, 16177377, 16190721, 16184169, 16184170, 16184171, 16184174, 16176381, 16176382, 16188747, 16188748, 16123236, 16199367, 16199368, 16199369, 16205650, 16205651, 16176365, 16176366, 16173019, 16163816, 16163818, 16191702, 16188734, 16188738, 16205201, 16204115, 16177171, 16177172, 16177173, 16177174, 16177176, 16177177, 16177179, 16177180, 16176414, 16176417, 16176418, 16176419, 16176420, 16176421, 16176432, 16171270, 16171271, 16201109, 16201110, 16198649, 16201105, 16116744, 16149564, 16149566, 16177169, 16177170, 16177175, 16189312, 16203229, 16203231, 16205158, 16173055, 16173056, 16172999, 16173000, 16205101, 16205102, 16095505, 16095508, 16168978, 16205277, 16188749, 16188751, 16188755, 16188756, 16188759, 16177178, 16204852, 16201104, 16182591, 16133700, 16203228, 16198623, 16191648, 16191653, 16171767, 16195516, 16205196, 16205197, 16205223, 16205224, 16204080, 16204082, 16204085, 16204086, 16178215, 16204087, 16204088, 16204090, 16204111, 16176416, 16205652, 16211243, 16211244, 16211245, 16201124, 16205161, 16207798, 16207799, 16178185, 16178186, 16204853, 16196921, 16204854, 16204855, 16203241, 15974555, 16205155, 16043788, 16207801, 16207803, 16182592, 16182593, 16182597, 16182602, 16182604, 16182613, 16182618, 16183926, 16205109, 16205111, 16192952, 16179760, 16198605, 16198606, 16198614, 16204144, 16211240, 16211241, 16211242, 16200052, 16200053, 16200054, 16205154, 16205159, 16207806, 16207808, 16131180, 16205123, 16191887, 16191888, 16191889, 16191890, 16191891, 16191892, 16191893, 16191894, 16191895, 16191896, 16191897, 16191898, 16191899, 16191900, 16191901, 16191902, 16193818, 16185813, 16193819, 16193821, 16193829, 16193830, 16193831, 16214368, 16214369, 16214370, 16203253, 16203254, 16203255, 16176377, 16211219, 16211220, 16211221, 16211223, 16211224, 16211225, 16211226, 16211227, 16211228, 16211229, 16211230, 16211231, 16211232, 16211233, 16211234, 16211235, 16211236, 16211237, 16211239, 16211255, 16211256, 16199978, 16189314, 16204560, 16205157, 16207810, 16207812, 16188752, 16188753, 16188754, 16188758, 16188760, 16177865, 16188762, 16188763, 16211246, 16211247, 16211249, 16211250, 16211251, 16207690, 16204157, 16204158, 16196918, 16204159, 16205153, 16154783, 16208475, 16208480, 16208484, 16208714, 16208740, 16203269, 16203270, 16203274, 16203275, 16203276, 16176433, 16201176, 16201177, 16201178, 16201179, 16161646, 16161647, 16161648, 16161649, 16161650, 16154392, 16161651, 16161652, 16161653, 16161654, 16154440, 16184198, 16184200, 16184890, 16184892, 16186503, 16186504, 16186505, 16184135, 16184136, 16196131, 16195510, 16195511, 16195513, 16178182, 16178183, 16204083, 16204084, 16204089, 16204091, 16204107, 16204164, 16179755, 16204165, 16204166, 16204168, 16204169, 16196919, 16196920, 16204170, 16204171, 16147350, 16147351, 16147352, 16147336, 16211252, 16211253, 16211254, 16210349, 16211211, 16211212, 16211213, 16207872, 16207874, 16182599, 16205117, 16196883, 16196885, 16196887, 16196890, 16196892, 16196893, 16196894, 16196895, 16196896, 16196897, 16196898, 16196899, 16177182, 16208477, 16209824, 16194908, 16194909, 16199355, 16208454, 16214381, 16122134, 16207877, 16207879, 16207881, 16207884, 16189305, 16200055, 16200056, 16200058, 16200059, 16196202, 16196203, 16208713, 16205188, 16205189, 16205280, 16205380, 16205381, 16199356, 16211259, 16211203, 16180831, 16173009, 16180830, 16180825, 16173010, 16214382, 16136905, 16136906, 16196913, 16196914, 16154635, 16154636, 16154638, 16154639, 16154640, 16154641, 16154642, 16154643, 16154644, 16154645, 16191886, 16196865, 16178173, 16178175, 16211261, 16213426, 16212794, 16212795, 16204856, 16204857, 16204858, 16163345, 16122683, 16203262, 16192935, 16205285, 16204167, 16177185, 16177186, 16177187, 16177370, 16177375, 16177188, 16210347, 16211201, 16194921, 16205670, 16205671, 16211560, 16211561, 16204162, 16199981, 16205690, 16205691, 16205191, 16205204, 16205221, 16205152, 16205164, 16197450, 16197451, 16197454, 16197456, 16197457, 16173003, 16180821, 16180820, 16180817, 16180815, 16182456, 16123660, 16123661, 16095506, 16213233, 16198652, 16198653, 16198654, 16198655, 16198656, 16198657, 16198658, 16198659, 16198660, 16102387, 16205192, 16215580, 16205227, 16205281, 16205282, 16205283, 16205284, 16205382, 16205383, 16205385, 16208452, 16188761, 16208453, 16169026, 16169027, 16167686, 16193895, 16193898, 16185816, 16193901, 16193902, 16193903, 16193906, 16193907, 16193908, 16193909, 16214380, 16214345, 16212349, 16199332, 16189296, 16205253, 16200060, 16201082, 16184151, 16193002, 16193820, 16193822, 16204145, 16204146, 16204148, 16213502, 16213503, 16211248, 16211262, 16173054, 16180829, 16180812, 16180810, 16180807, 16182454, 16182584, 16182606, 16182608, 16205105, 16205118, 16205122, 16204160, 16204161, 16163830, 16199360, 16199361, 16199362, 16199363, 16199364, 16199365, 16203243, 16203250, 16196132, 16196133, 16196134, 16196135, 16196137, 16215416, 16180771, 16180776, 16176460, 16190717, 16211258, 16211260, 16212796, 16207684, 16205218, 16205258, 16207692, 16207693, 16207694, 16207696, 16203256, 16189303, 16218444, 16218445, 16211521, 16211522, 16211523, 16211524, 16211526, 16211527, 16192545, 16211528, 16211529, 16192544, 16215420, 16073439, 16073441, 16073444, 16075275, 16075276, 16075277, 16075279, 16201106, 16205173, 16205175, 16205178, 16205179, 16205180, 16205181, 16205182, 16208715, 16214675, 16053308, 16214344, 16171755, 16171757, 16171759, 16171762, 16135309, 16135310, 16135312, 16205176, 16205177, 16188743, 16196206, 16196207, 16203283, 16201090, 16201095, 16201096, 16211490, 16211518, 16211519, 16211520, 16214346, 16208983, 16213425, 16211512, 16211513, 16211514, 16211515, 16205170, 16205172, 16196204, 16196205, 16196208, 16196209, 16196215, 16208719, 16219576, 16103916, 16214377, 16214379, 16203251, 16192548, 16196139, 16196140, 16196141, 16196143, 16196144, 16196146, 16196147, 16196148, 16196149, 16196150, 16195508, 16195509, 16195512, 16195517, 16216451, 16216452, 16203278, 16203279, 16203281, 16215590, 16216465, 16126827, 16126828, 16208718, 16208722, 16208730, 16204130, 16204131, 16176425, 16163343, 16163344, 16205104, 16163346, 16205208, 16205216, 16210245, 16210250, 16210252, 16213456, 16213457, 16213900, 16217600, 16217601, 16217603, 16219091, 16211491, 16211492, 16211493, 16211494, 16211495, 16211496, 16211497, 16211499, 16211500, 16215508, 16215509, 16205184, 16205183, 16208466, 16208478, 16208486, 16208724, 16208729, 16208732, 16208733, 16209833, 16209839, 16210237, 16210246, 16216405, 16216407, 16216408, 16216411, 16203277, 16203282, 16220096, 16220099, 16220103, 16220105, 16220113, 16220107, 16220108, 16220110, 16205672, 16215507, 16205106, 16205107, 16205108, 16205110, 16205113, 16205115, 16205120, 16205121, 16205661, 16205663, 16205664, 16217557, 16217558, 16217559, 16217561, 16217563, 16217565, 16217567, 16196901, 16190718, 16164729, 16165154, 16196200, 16196201, 16196212, 16215548, 16215556, 16215557, 16185793, 16183924, 16219536, 16219537, 16219538, 16217585, 16217587, 16217590, 16217593, 16191680, 16183925, 16207678, 16207679, 16207680, 16207682, 16204143, 16176379, 16176380, 16176424, 16116319, 16213893, 16208742, 16209838, 16136848, 16136849, 16136850, 16163237, 16204136, 16204138, 16204141, 16204140, 16196880, 16196878, 16196879, 16090323, 16095507, 16177183, 16165905, 16211210, 16217576, 16217577, 16217522, 16217523, 16217552, 16217553, 16217554, 16213230, 16198664, 16215274, 16215427, 16213251, 16203280, 16217578, 16217580, 16217581, 16217582, 16217583, 16217584, 16217618, 16217626, 16217627, 16218429, 16218431, 16218440, 16220097, 16220098, 16220101, 16220109, 16176423, 16163341, 16176426, 16176427, 16196127, 16196129, 16195519, 16195521, 16195522, 16195524, 16195525, 16195526, 16216454, 16215549, 16215552, 16215553, 16208727, 16208736, 16209835, 16205203, 16210238, 16211510, 16211511, 16205396, 16205397, 16205398, 16205400, 16205401, 16205402, 16205403, 16195498, 16195499, 16195500, 16195501, 16163323, 16163324, 16163325, 16195495, 16195496, 16219539, 16216466, 16216467, 16221118, 16213224, 16213225, 16213226, 16160081, 16196211, 16150304, 16154760, 16150310, 16150311, 16150313, 16215425, 16215429, 16214347, 16214348, 16214351, 16214352, 16214353, 16214354, 16193893, 16193894, 16193900, 16193904, 16193905, 16193911, 16147591, 16147592, 16192556, 16193963, 16193965, 16193967, 16193969, 16193971, 16193972, 16217586, 16217588, 16217592, 16217595, 16217596, 16217598, 16211202, 16217616, 16217619, 16217621, 16217625, 16221153, 16220106, 16220111, 16220112, 16217189, 16217195, 16219173, 16219176, 16219178, 16219179, 16195518, 16198619, 16216457, 16210239, 16196121, 16217617, 16218430, 16218438, 16218442, 16211217, 16205141, 16205142, 16205143, 16205144, 16205145, 16205146, 16205147, 16205148, 16211557, 16211559, 16214341, 16154746, 16205689, 16214663, 16208471, 16208474, 16222578, 16213432, 16213433, 16213435, 16213436, 16213437, 16213442, 16213445, 16213446, 16213449, 16176431, 16167700, 16167702, 16188701, 16217624, 16218434, 16185795, 16185794, 16191711, 16211489, 16211556, 16211558, 16211498, 16041388, 16041514, 16041515, 16041516, 16041651, 16211507, 16211508, 16211509, 16205662, 16205666, 16205667, 16217560, 16217562, 16217564, 16217566, 16217568, 16217569, 16205406, 16208470, 16216453, 16186581, 16214355, 16214356, 16214357, 16214358, 16214359, 16214360, 16214362, 16196923, 16196924, 16196925, 16196926, 16205098, 16205099, 16205100, 16201088, 16201089, 16221276, 16214365, 16214367, 16201117, 16201119, 16201120, 16214667, 16214668, 16214670, 16214671, 16214673, 16212391, 16214374, 16221145, 16218453, 16218454, 16204155, 16204156, 16221150, 16221151, 16221158, 16221160, 16221526, 16221527, 16221528, 16221529, 16221531, 16221533, 16221535, 16221536, 16221537, 16213228, 16147338, 16211531, 16214664, 16208472, 16208473, 16189316, 16189318, 16204150, 16189319, 16189320, 16189322, 16204151, 16204152, 16204153, 16204154, 16205193, 16205194, 16205195, 16205217, 16205219, 16205222, 16211562, 16211563, 16211564, 16213438, 16213447, 16205694, 16153577, 16205696, 16205697, 16214401, 16214402, 16211554, 16211555, 16199357, 16204094, 16215592, 16215593, 16221808, 16205660, 16186587, 16217203, 16154650, 16154742, 16154747, 16189307, 16214349, 16220076, 16220077, 16220078, 16221530, 16221142, 16219535, 16215589, 16199970, 16199971, 16199972, 16199974, 16199976, 16214672, 16225002, 16207677, 16219194, 16192937, 16192938, 16192939, 16192940, 16192943, 16208459, 16182532, 16224460, 16219577, 16219579, 16219580, 16219581, 16219582, 16219583, 16219584, 16219585, 16221866, 16192560, 16221286, 16221287, 16221290, 16225402, 16221303, 16221304, 16221305, 16221314, 16192942, 16216449, 16216450, 16215560, 16215562, 16217597, 16217599, 16219534, 16220638, 16220639, 16220640, 16220641, 16201130, 16201131, 16201132, 16201133, 16201134, 16201135, 16201136, 16217623, 16217620, 16217622, 16218433, 16218435, 16218437, 16218439, 16218441, 15955579, 16207710, 16207729, 16207741, 16207744, 16207746, 16207750, 16221149, 16221155, 16221156, 16221159, 16221119, 16221120, 16182554, 16208985, 16208986, 16221822, 16189317, 16189321, 16144476, 16221804, 16216456, 16216458, 16004224, 16184179, 16196210, 16205190, 16218502, 16218503, 16219532, 16211238, 16211257, 16154741, 16154743, 16154744, 16154745, 16154750, 16207819, 16198686, 16198688, 16198689, 16198643, 16198645, 16198646, 16198647, 16198648, 16213430, 16213431, 16213440, 16213443, 16160089, 16160090, 16160091, 16213461, 16213466, 16196856, 16196857, 16218451, 16221809, 16219545, 16219546, 16063451, 16075308, 16216459, 16216461, 16215539, 16215551, 16163813, 16219533, 16208449, 16208450, 16208451, 16216429, 16216436, 16215585, 16219036, 16219040, 16219041, 16215591, 16216469, 16216478, 16216486, 16216488, 16217570, 16217571, 16216460, 16223923, 16217609, 16217524, 16218459, 16218460, 16218461, 16223410, 16223411, 16154649, 16154748, 16154749, 16203264, 16222567, 16217547, 16217549, 16207715, 16207718, 16207720, 16207734, 16207736, 16207745, 16207754, 16207755, 16205116, 16205665, 16191883, 16191885, 16198611, 16195523, 16216448, 16215537, 16215554, 16215558, 16215559, 16215565, 16217527, 16217528, 16217530, 16220075, 16219084, 16219085, 16214412, 16224457, 16224458, 16224459, 16221877, 16178168, 16208746, 16208748, 16209820, 16209827, 16220058, 16220059, 16220060, 16220062, 16217533, 16217534, 16215597, 16217602, 16211481, 16211482, 16211483, 16211484, 16211486, 16211487, 16221833, 16221834, 16221835, 16221836, 16221837, 16221838, 16221839, 16221840, 16221841, 16221842, 16221843, 16221844, 16221845, 16221846, 16221847, 16221848, 16205103, 16172937, 16178184, 16178187, 16165761, 16165772, 16165773, 16165774, 16165775, 16184900, 16219037, 16219038, 16219045, 16219049, 16213460, 16216474, 16216476, 16216480, 16216485, 16216487, 16117996, 16118774, 16118775, 16217183, 16217184, 16217185, 16217186, 16217187, 16217188, 16217190, 16217191, 16217192, 16217193, 16217194, 16217196, 16217197, 16217198, 16217199, 16217200, 16217201, 16217202, 16217204, 16217205, 16217206, 16217207, 16217208, 16217209, 16217210, 16217211, 16217212, 16217213, 16217214, 16219174, 16219175, 16219177, 16219555, 16219556, 16219558, 16219559, 16219560, 16219566, 16219567, 16219571, 16219572, 16219573, 16226381, 16226382, 16226383, 16226385, 16226386, 16226387, 16226389, 16226392, 16226393, 16226394, 16226632, 16226652, 16226653, 16226658, 16226659, 16221503, 16221504, 16221525, 16221258, 16221262, 16222579, 16222580, 16220642, 16220643, 16217604, 16216484, 16216490, 16216491, 16216492, 16216493, 16218432, 16218447, 16218449, 16223107, 16223108, 16207752, 16207753, 16208461, 16208463, 16208464, 16220115, 16220116, 16220118, 16220120, 16220123, 16214342, 16223934, 16223935, 16223936, 16223937, 16221810, 16221813, 16221817, 16221818, 16221819, 16221820, 16221821, 16217526, 16217529, 16217531, 16220061, 16220063, 16220064, 16221807, 16228172, 16223414, 16223421, 16220100, 16220102, 16220104, 16211485, 16211488, 16223109, 16221811, 16221812, 16221816, 16127312, 16217574, 16217575, 16217594, 16219529, 16219530, 16219531, 16226664, 16208738, 16208745, 16209812, 16209819, 16209822, 16209825, 16139221, 16221162, 16221165, 16221249, 16221252, 16221259, 16221260, 16213459, 16223417, 16223418, 16223422, 16227143, 16226310, 16226311, 16226313, 16217551, 16222575, 16226644, 16226645, 16226646, 16226647, 16226648, 16226583, 16226584, 16226585, 16226586, 16226587, 16226588, 16226590, 16226591, 16226594, 16226595, 16226596, 16226597, 16226599, 16226600, 16226601, 16226602, 16226603, 16226606, 16226607, 16219565, 16219569, 16122601, 16223120, 16223121, 16223122, 16223123, 16184901, 16190708, 16184922, 16184924, 16184925, 16213450, 16221263, 16221849, 16221850, 16221851, 16221852, 16221853, 16176462, 16180778, 16180796, 16226309, 16219195, 16219574, 16219575, 16221147, 16221148, 16221152, 16221154, 16221157, 16226312, 16177197, 16220114, 16220117, 16220121, 16220122, 16217138, 16215594, 16222574, 16222576, 16222548, 16222549, 16222550, 16222551, 16222552, 16222553, 16219158, 16219166, 16219187, 16178166, 16227361, 16217555, 16217556, 16117961, 16226614, 16226619, 16226620, 16226654, 16226655, 16226656, 16226657, 16226662, 16215563, 16228116, 16228118, 16221893, 16221806, 16204850, 16204851, 16221251, 16221255, 16221257, 16221261, 16222581, 16223942, 16223943, 16223944, 16223945, 16223946, 16223947, 16223949, 16223950, 16223951, 16223953, 16223954, 16223955, 16213458, 16227136, 16004223, 16221520, 16205405, 16199366, 16203268, 16208467, 16208735, 16208743, 16209814, 16209816, 16209823, 16209826, 16228117, 16228528, 16219167, 16219168, 16219171, 16219186, 16226308, 16223415, 16223416, 16220653, 16220654, 16220655, 16220657, 16223406, 16135300, 16160082, 16220597, 16220598, 16209837, 16217503, 16221895, 16227127, 16222582, 16223413, 16223956, 16208465, 16215512, 16215513, 16215514, 16215516, 16215518, 16215519, 16215520, 16215521, 16215522, 16215523, 16215540, 16215541, 16215543, 16215547, 16226589, 16226592, 16226593, 16226598, 16226604, 16226605, 16226608, 16226609, 16221281, 16221282, 16221283, 16221284, 16192559, 16221288, 16221289, 16221291, 16221292, 16221293, 16221294, 16221295, 16221296, 16221298, 16221299, 16221300, 16221301, 16221302, 16221306, 16221307, 16221308, 16221309, 16221310, 16221311, 16221312, 16221313, 16165151, 16221872, 16221873, 16221874, 16221875, 16221876, 16221878, 16221879, 16221881, 16221883, 16221884, 16221886, 16221887, 16221888, 16217573, 16215555, 16205276, 16205278, 16215436, 16215437, 16215438, 16190707, 16184927, 16221823, 16221824, 16221825, 16221826, 16145590, 16217589, 16217591, 16214698, 16221161, 16221163, 16221854, 16221855, 16221856, 16221857, 16221858, 16221859, 16221862, 16221864, 16191644, 16192558, 16216470, 16216477, 16216481, 16216482, 16216483, 16221266, 16215485, 16215489, 16227082, 16229824, 16223111, 16223112, 16223113, 16223114, 16223115, 16223116, 16223117, 16223118, 16223119, 16229975, 16229983, 16229993, 16229995, 16229998, 16230000, 16230001, 16135307, 16226615, 16226616, 16226621, 16226665, 16226666, 16209831, 16228527, 16211533, 16220067, 16215440, 16215442, 16215449, 16215455, 16215459, 16215463, 16222583, 16216472, 16220656, 16221502, 16230567, 16207722, 16207725, 16207731, 16207739, 16207749, 16208462, 16211501, 16211502, 16211503, 16211504, 16211505, 16211506, 16230575, 16230579, 16230585, 16230588, 16230597, 16231386, 16231388, 16209829, 16221166, 16154525, 16203287, 16221264, 16221265, 16223125, 16218496, 16123244, 16123245, 16123246, 16123247, 16123248, 16226613, 16226617, 16226618, 16226650, 16226651, 16226668, 16219039, 16219050, 16219051, 16214393, 16214394, 16214395, 16216471, 16216473, 16216479, 16216494, 16226641, 16223126, 16223127, 16221146, 16219542, 16219544, 16219547, 16219548, 16219550, 16226325, 16226331, 16226334, 16226339, 16226340, 16226341, 16214422, 16214423, 16214424, 16214425, 16214426, 16214660, 16214661, 16221532, 16221534, 16221894, 16223928, 16221253, 16221254, 16215595, 16226643, 16221521, 16230576, 16230578, 16230584, 16230587, 16230590, 16230591, 16230596, 16231389, 16229989, 16230327, 16230329, 16230331, 16230332, 16230334, 16230343, 16226660, 16227056, 16227059, 16227060, 16227061, 16221164, 16221250, 16217179, 16217180, 16217181, 16219117, 16236188, 16223464, 16223467, 16218466, 16218489, 16218491, 16218492, 16226667, 16227057, 16236647, 16221167, 16221256, 16229808, 16229823, 16229974, 16229976, 16229979, 16229987, 16229990, 16230328, 16230335, 16230336, 16230337, 16230340, 16230341, 16230342, 16230344, 16230345, 16226611, 16226612, 16226649, 16226663, 16227058, 16205235, 16205238, 16205242, 16205246, 16205264, 16205268, 16205273, 16205275, 16123239, 16123241, 16177194, 16227369, 16212800, 16212801, 16212802, 16212803, 16219042, 16219044, 16219046, 16219047, 16228173, 16223419, 16223420, 16223465, 16223466, 16229813, 16229815, 16229817, 16222568, 16231404, 16236642, 16236645, 16214679, 16214681, 16214683, 16214684, 16214685, 16214687, 16214688, 16214690, 16214691, 16214692, 16214693, 16214694, 16214695, 16214696, 16214697, 16216424, 16216427, 16216431, 16222571, 16222573, 16222585, 16222586, 16224456, 16223104, 16228120, 16228121, 16228122, 16189299, 16189301, 16231358, 16214403, 16214405, 16214407, 16214408, 16214409, 16214411, 16214415, 16214416, 16214417, 16211517, 16171761, 16211532, 16211534, 16211537, 16229195, 16229199, 16214364, 16220644, 16176225, 16226332, 16226338, 16221505, 16221506, 16221507, 16221508, 16221510, 16221511, 16221512, 16221513, 16221514, 16221515, 16205202, 16210247, 16210248, 16210249, 16211535, 16211536, 16211538, 16211539, 16229194, 16221519, 16213462, 16213463, 16213464, 16213467, 16239364, 16223472, 16230577, 16230580, 16230581, 16230582, 16230583, 16230592, 16230593, 16230594, 16230595, 16230598, 16230599, 16230600, 16231385, 16231387, 16231390, 16122238, 16236644, 16236650, 16139205, 16236132, 16048030, 16223468, 16223470, 16223471, 16231459, 16231460, 16231461, 16231462, 16232262, 16232263, 16232264, 16232267, 16234840, 16234841, 16234842, 16221509, 16205668, 16236643, 16219957, 16222569, 16222570, 16222572, 16223102, 16223103, 16211540, 16211541, 16211542, 16211543, 16211544, 16211545, 16211546, 16211548, 16211549, 16211550, 16229191, 16229192, 16229193, 16229196, 16229197, 16229198, 16223106, 16237311, 16237312, 16223930, 16223931, 16223932, 16223933, 16223469, 16223425, 16223426, 16223427, 16223429, 16228902, 16228903, 16219035, 16208468, 16208469, 16196927, 16196928, 16196929, 16196930, 16196931, 16234515, 16212781, 16221134, 16221137, 16221138, 16221141, 16215432, 16226317, 16226318, 16226319, 16226320, 16226322, 16226324, 16226326, 16226327, 16226329, 16226330, 16226333, 16226337, 16226342, 16226344, 16236641, 16236649, 16226305, 16226306, 16226307, 16150294, 16150295, 16213465, 16220645, 16220646, 16220647, 16220648, 16220649, 16220650, 16220651, 16220652, 16227375, 16228007, 16228009, 16228010, 16223139, 16223140, 16223141, 16223142, 16217613, 16217615, 16091218, 16091219, 16091221, 16091856, 16218485, 16218487, 16218494, 16218495, 16208476, 16208481, 16208483, 16208485, 16208487, 16208707, 16208711, 16208712, 16236648, 16228170, 16154590, 16219159, 16219160, 16219172, 16219185, 16220082, 16223423, 16223424, 16236157, 16236173, 16236174, 16236178, 16239926, 16241408, 16205391, 16205392, 16205394, 16231464, 16232259, 16234832, 16234833, 16234834, 16234835, 16234836, 16234844, 16218477, 16218478, 16218479, 16218480, 16218498, 16218500, 16231397, 16231398, 16231401, 16231402, 16231405, 16228169, 16228174, 16228175, 16228176, 16228177, 16228178, 16228179, 16228180, 16228181, 16228186, 16228188, 16228189, 16228194, 16228195, 16228199, 16228201, 16228202, 16228205, 16228907, 16228913, 16228167, 16228168, 16227099, 16227100, 16227103, 16227104, 16227105, 16227107, 16227109, 16227111, 16234514, 16213147, 16223124, 16218446, 16214677, 16228915, 16207751, 16236719, 16236720, 16230589, 16237310, 16227146, 16227147, 16208479, 16208716, 16208725, 16220071, 16220072, 16220073, 16220074, 16227101, 16229773, 16219161, 16219170, 16227376, 16228005, 16228006, 16228008, 16228011, 16176438, 16236171, 16236172, 16236176, 16236175, 16236177, 16236179, 16236180, 16236654, 16236655, 16236657, 16236658, 16231463, 16232260, 16232261, 16232265, 16232266, 16234838, 16234839, 16205681, 16192547, 16203242, 16228119, 16219958, 16219964, 16219965, 16219969, 16219970, 16219971, 16228906, 16228908, 16214384, 16214392, 16214396, 16214397, 16223143, 16217521, 16217525, 16214414, 16221880, 16217504, 16217506, 16217505, 16221896, 16229189, 16227144, 16230403, 16230404, 16228877, 16228886, 16231378, 16231379, 16231380, 16233794, 16236697, 16236701, 16239953, 16239954, 16091849, 16091850, 16164694, 16142157, 16061940, 16056732, 16056733, 16057170, 16057174, 16241572, 16236670, 16241573, 16241574, 16236671, 16231430, 16231431, 16237336, 16176435, 16231360, 16231361, 16231362, 16231364, 16231365, 16236651, 16236653, 16231366, 16231367, 16231369, 16221827, 16221828, 16227385, 16211551, 16211552, 16199371, 16194930, 16231381, 16227102, 16231444, 16230389, 16230540, 16230546, 16230551, 16236164, 16236646, 16234516, 16234517, 16233795, 16241410, 16240038, 16240039, 16154355, 16154356, 16207795, 16091216, 16091217, 16164691, 16182545, 16142155, 16066118, 16142156, 16207797, 16164709, 16229782, 16229784, 16229786, 16229787, 16229789, 16229793, 16231359, 16231455, 16221136, 16215525, 16215526, 16226314, 16226315, 16226321, 16226323, 16226343, 16226345, 16226346, 16226347, 16084300, 16243648, 16243650, 16243652, 16243654, 16243659, 16243662, 16243664, 16243666, 16243668, 16243670, 16243672, 16243674, 16243614, 16243682, 16244026, 16237338, 16231396, 16231399, 16231403, 16243337, 16243339, 16219961, 16221143, 16221144, 16227378, 16227379, 16227384, 16227386, 16231376, 16211547, 16228881, 16228884, 16228888, 16234872, 16234873, 16234874, 16223941, 16214383, 16214386, 16214387, 16214389, 16214390, 16217605, 16217606, 16217607, 16236148, 16236149, 16236150, 16236151, 16236152, 16236153, 16236154, 16236155, 16236156, 16236158, 16236160, 16239517, 16223136, 16223137, 16223138, 16223148, 16214413, 16227145, 16218450, 16233796, 16239266, 16239273, 16239274, 16239275, 16239276, 16239277, 16239278, 16239279, 16239281, 16239282, 16239283, 16239286, 16239288, 16231456, 16241419, 16241421, 16241566, 16236660, 16236661, 16236662, 16236663, 16238709, 16241567, 16241568, 16236664, 16236666, 16236668, 16236669, 16240413, 16240414, 16241571, 16241575, 16219563, 16219564, 16226628, 16164742, 16237332, 16228931, 16228932, 16229239, 16228917, 16228918, 16219962, 16231368, 16231370, 16227377, 16227380, 16231375, 16228183, 16228185, 16228187, 16228193, 16228196, 16228197, 16228198, 16228203, 16228204, 16228887, 16164684, 16163390, 16214366, 16227110, 16229255, 16231440, 16231445, 16244619, 16244625, 16230395, 16230543, 16230568, 16240034, 16240035, 16240036, 16240037, 16240040, 16243656, 16243675, 16243678, 16243679, 16243687, 16244024, 16244031, 16244038, 16241430, 16241565, 16228889, 16228891, 16239703, 16237334, 16243333, 16243335, 16243342, 16154524, 16240502, 16240508, 16243331, 16227106, 16227108, 16223957, 16223958, 16207713, 16221277, 16221278, 16221279, 16221280, 16221285, 16245022, 16245023, 16227381, 16227382, 16227383, 16231372, 16231374, 16231377, 16221524, 16221297, 16220119, 16242630, 16242631, 16230385, 16230386, 16230387, 16230393, 16230394, 16230544, 16236713, 16236717, 16240489, 16241658, 16241661, 16237340, 16230376, 16230378, 16231400, 16208455, 16213240, 16215573, 16210234, 16231371, 16231373, 16227387, 16229801, 16229803, 16229804, 16229806, 16229807, 16240493, 16240501, 16230388, 16230548, 16240421, 16154357, 16220083, 16243325, 16239969, 16239970, 16239971, 16177189, 16179784, 16210345, 16241417, 16219551, 16219552, 16221889, 16221890, 16221892, 16234843, 16234845, 16234846, 16234847, 16234848, 16234850, 16234851, 16234852, 16234853, 16234854, 16216438, 16234855, 16234856, 16228874, 16228879, 16228880, 16228890, 16234512, 16231438, 16231439, 16231441, 16231442, 16231443, 16213468, 16213471, 16213472, 16213473, 16213474, 16213475, 16213476, 16213477, 16239267, 16239269, 16239270, 16239271, 16239272, 16239284, 16239285, 16219549, 16219553, 16219554, 16221126, 16221127, 16221139, 16221140, 16215434, 16215528, 16222577, 16239699, 16232258, 16244622, 16244624, 16244626, 16244628, 16245329, 16245001, 16245002, 16243644, 16243646, 16243616, 16243681, 16243683, 16244023, 16244025, 16244027, 16244029, 16244030, 16244032, 16244033, 16244035, 16244036, 16244037, 16244039, 16219959, 16219960, 16219963, 16219966, 16219967, 16219968, 16219973, 16219975, 16219977, 16219979, 16219981, 16229850, 16229852, 16248880, 16248882, 16234510, 16217508, 16240511, 16228526, 16228882, 16228904, 16220618, 16220619, 16220620, 16220621, 16220622, 16219071, 16219072, 16219074, 16219075, 16219076, 16219082, 16223134, 16223135, 16223145, 16223151, 16223152, 16223155, 16231382, 16231383, 16211207, 16211208, 16211209, 16248874, 16248878, 16248896, 16220085, 16243362, 16243374, 16243377, 16243379, 16244694, 16244695, 16244696, 16236165, 16236166, 16236168, 16236170, 16215503, 16215505, 16215506, 16063112, 16063113, 16231433, 16242674, 16243676, 16243329, 16223938, 16223939, 16223940, 16228859, 16228860, 16228905, 16228914, 16154522, 16154528, 16163315, 16223105, 16243999, 16244005, 16244008, 16244009, 16244010, 16241917, 16241918, 16241922, 16241924, 16241925, 16241926, 16241927, 16241943, 16241944, 16241945, 16241947, 16241948, 16221830, 16221831, 16221832, 16240422, 16223462, 16223463, 16164690, 16164692, 16091851, 16091853, 16243622, 16243626, 16243618, 16208457, 16208460, 16205198, 16213236, 16210221, 16213242, 16213243, 16213246, 16210222, 16210240, 16210243, 16210253, 16223428, 16236186, 16236190, 16236191, 16236192, 16221516, 16222559, 16222560, 16222561, 16222562, 16222563, 16222564, 16222565, 16222566, 16243361, 16243370, 16240494, 16240505, 16226633, 16244777, 16241946, 16230392, 16218455, 16218457, 16229856, 16228919, 16231453, 16231454, 16236718, 16236721, 16236722, 16244999, 16245000, 16248804, 16248806, 16240415, 16240416, 16240417, 16241576, 16232467, 16232468, 16232469, 16232470, 16232471, 16232472, 16232474, 16232475, 16232476, 16232477, 16232478, 16231466, 16198612, 16198613, 16203266, 16243313, 16248875, 16234857, 16234859, 16234860, 16234861, 16234863, 16234865, 16234867, 16234868, 16234869, 16234870, 16220084, 16239256, 16220088, 16228862, 16243372, 16243373, 16243378, 16243381, 16239998, 16245356, 16245357, 16247505, 16247506, 16245358, 16245360, 16229847, 16229848, 16234530, 16227124, 16241656, 16237318, 16237319, 16237320, 16237321, 16237323, 16240441, 16240442, 16229790, 16240443, 16240444, 16229791,

More Code Info

16240445, 16240446, 16188745, 16154780, 16245014, 16230349, 16243368, 16240023, 16240024, 16240025, 16240027, 16240031, 16227098, 16214398, 16182533, 16236140, 16236142, 16236144, 16236146, 16236147, 16237313, 16219162, 16219164, 16219169, 16219181, 16219182, 16219183, 16219189, 16223130, 16223131, 16223153, 16223154, 16214404, 16245771, 16230372, 16244776, 16249409, 16205390, 16245324, 16245325, 16245326, 16245327, 16245331, 16236638, 16236639, 16192552, 16192553, 16208456, 16215578, 16192555, 16210224, 16222584, 16245469, 16242558, 16245769, 16245770, 16249639, 16247530, 16247532, 16227148, 16221123, 16221124, 16221125, 16221128, 16215524, 16215529, 16245347, 16245349, 16245351, 16245353, 16182525, 16247503, 16247504, 16234858, 16234864, 16220090, 16228182, 16228184, 16243360, 16243363, 16243364, 16243366, 16243367, 16243371, 16243375, 16243376, 16243380, 16243382, 16243383, 16243385, 16244314, 16244316, 16244317, 16244318, 16245778, 16245779, 16245780, 16223128, 16223129, 16223133, 16248741, 16247531, 16240425, 16247533, 16247534, 16219543, 16243283, 16243284, 16243285, 16243286, 16245025, 16245026, 16243288, 16241569, 16241570, 16243290, 16243291, 16245039, 16243293, 16243294, 16243608, 16243609, 16245040, 16198616, 16198617, 16192947, 16192951, 16248872, 16243344, 16245012, 16240043, 16240044, 16197428, 16244657, 16244659, 16197429, 16197430, 16197431, 16197432, 16197433, 16197434, 16227114, 16227116, 16241578, 16241580, 16231450, 16244623, 16229819, 16236665, 16220079, 16239701, 16239960, 16239961, 16239962, 16239963, 16239964, 16239965, 16239966, 16245792, 16048024, 16048040, 16154389, 16245008, 16234875, 16234876, 16234877, 16245010, 16221498, 16229849, 16229851, 16229853, 16229859, 16229861, 16229863, 16229866, 16231468, 16231469, 16231470, 16231471, 16239681, 16239682, 16239689, 16239690, 16239692, 16247539, 16247540, 16231434, 16241862, 16241863, 16241915, 16241907, 16245366, 16245367, 16245372, 16241908, 16241919, 16221622, 16243365, 16243369, 16247526, 16247528, 16223404, 16239260, 16239261, 16239265, 16236136, 16244311, 16242635, 16243640, 16242636, 16249411, 16199331, 16249427, 16249431, 16245419, 16240500, 16240504, 16241647, 16241649, 16223156, 16211218, 16241579, 16229774, 16229776, 16249656, 16249657, 16244190, 16244192, 16244193, 16244195, 16244198, 16244200, 16244201, 16241959, 16242597, 16242602, 16242606, 16242612, 16242619, 16242560, 16237322, 16237324, 16219515, 16219516, 16219517, 16219518, 16219519, 16219520, 16219521, 16219522, 16219523, 16219524, 16219525, 16219526, 16219527, 16219528, 16236185, 16236187, 16236189, 16228014, 16228015, 16228017, 16228021, 16228022, 16228024, 16228025, 16228027, 16228030, 16228210, 16228211, 16228212, 16228213, 16228214, 16258394, 16258399, 16253099, 16253100, 16249459, 16219540, 16219541, 16245021, 16239680, 16247519, 16247520, 16247521, 16243619, 16243624, 16192931, 16192932, 16236181, 16236182, 16236183, 16236184, 16223412, 16147323, 16247512, 16222554, 16249458, 16244629, 16244632, 16244654, 16247537, 16239683, 16249665, 16244490, 16244491, 16244493, 16244495, 16244496, 16244497, 16244498, 16221891, 16221885, 16186506, 16186507, 16186508, 16186509, 16186510, 16186511, 16186512, 16186513, 16186514, 16186515, 16187826, 16188317, 16188318, 16188320, 16188321, 16188323, 16188324, 16188350, 16188351, 16188352, 16126052, 16249452, 16247508, 16244487, 16244488, 16219163, 16219165, 16219180, 16219184, 16219190, 16219191, 16237315, 16237316, 16253257, 16242664, 16242666, 16242668, 16242672, 16242676, 16242677, 16243282, 16243632, 16213239, 16213241, 16213244, 16215570, 16215572, 16215574, 16215575, 16215576, 16215577, 16213247, 16205200, 16210225, 16210229, 16213249, 16260117, 16260118, 16260119, 16260120, 16260121, 16260123, 16231446, 16231447, 16231448, 16231449, 16241416, 16220089, 16248242, 16248247, 16248249, 16248251, 16248254, 16248256, 16248258, 16248300, 16248470, 16248478, 16248479, 16243680, 16242603, 16185806, 16147334, 16147335, 16127361, 16249433, 16249434, 16247489, 16247490, 16241920, 16228016, 16228018, 16228019, 16228020, 16228023, 16228028, 16228029, 16228207, 16228208, 16228209, 16239997, 16239999, 16257565, 16242670, 16244310, 16241906, 16244313, 16211568, 16244303, 16260127, 16211194, 16211195, 16211196, 16211197, 16211198, 16211199, 16211200, 16240026, 16207697, 16226634, 16226635, 16226636, 16227140, 16226637, 16226638, 16226639, 16226640, 16249461, 16230541, 16248710, 16244703, 16242596, 16241910, 16241911, 16259308, 16261356, 16227083, 16227084, 16227085, 16227086, 16227087, 16227088, 16227089, 16227090, 16227091, 16227092, 16227373, 16227374, 16231413, 16231414, 16231415, 16244015, 16262279, 16262280, 16239518, 16239538, 16239539, 16239540, 16239541, 16239543, 16223147, 16239546, 16239628, 16239634, 16239637, 16240041, 16243628, 16250946, 16250947, 16250948, 16250953, 16250954, 16250956, 16250958, 16250959, 16250960, 16250964, 16250966, 16252431, 16244302, 16215567, 16213245, 16215579, 16215587, 16215588, 16260125, 16249429, 16249430, 16254451, 16228013, 16228216, 16228218, 16228219, 16228221, 16247510, 16240029, 16213499, 16213501, 16227141, 16227142, 16244186, 16244188, 16253256, 15817326, 15817328, 16249432, 16250957, 16249435, 16249436, 16253097, 16253441, 16253442, 16223144, 16243639, 16218499, 16218501, 16204860, 16250943, 16250944, 16250949, 16250950, 16250951, 16250952, 16250955, 16250961, 16250963, 16250965, 16250967, 16250968, 16250970, 16250971, 16250972, 16250973, 16250974, 16250975, 16250976, 16250977, 16250978, 16250979, 16250980, 16250981, 16250982, 16252424, 16252425, 16252429, 16260124, 16260126, 16242628, 16217507, 16242629, 16247491, 16252486, 16249641, 16249642, 16249643, 16249644, 16249646, 16249650, 16227371, 16227372, 16254956, 16247525, 16247538, 16247527, 16247529, 16240427, 16240429, 16247535, 16247536, 16118773, 16118776, 16114690, 16114696, 16114699, 16242584, 16242585, 16242586, 16242587, 16241950, 16241951, 16241953, 16241954, 16241957, 16260132, 16244324, 16244483, 16244484, 16244485, 16244486, 16244489, 16241949, 16230390, 16230391, 16230545, 16229833, 16252464, 16252467, 16262749, 16252470, 16247493, 16247495, 16247498, 16198665, 16245077, 16245083, 16245087, 16234866, 16244001, 16244017, 16244018, 16244021, 16243610, 16243611, 16243612, 16254471, 16253103, 16246252, 16246257, 16246259, 16246262, 16246278, 16246282, 16263253, 16263254, 16264489, 16264490, 16244315, 16247492, 16247496, 16247497, 16247501, 16243630, 16243631, 16249443, 16259304, 16261360, 16261362, 16220087, 16240491, 16240492, 16240506, 16240507, 16266068, 16254472, 16245731, 16259305, 16260128, 16261361, 16261373, 16254466, 16245064, 16245070, 16245072, 16245079, 16254453, 16254455, 16254456, 16254457, 16254459, 16228220, 16248885, 16248898, 16248901, 16223132, 16223502, 16223504, 16223505, 16223506, 16223507, 16223920, 16268539, 16240431, 16240434, 16264502, 16264503, 16241411, 16241412, 16241413, 16241415, 16248218, 16248219, 16248220, 16248222, 16248223, 16248224, 16248225, 16253447, 16253622, 16254038, 16254043, 16248234, 16248237, 16248261, 16248297, 16248299, 16248301, 16248465, 16248466, 16248469, 16248472, 16248473, 16248475, 16248477, 16248480, 16248481, 16261371, 16223409, 16253435, 16240419, 16249640, 16249648, 16249652, 16231412, 16231416, 16250438, 16246267, 16244482, 16254479, 16262754, 16262756, 16239402, 16219060, 16219065, 16208458, 16130562, 16223929, 16245420, 16236714, 16239693, 16265923, 16248215, 16253450, 16253451, 16253616, 16253618, 16253619, 16226335, 16254037, 16254040, 16254041, 16254047, 16254050, 16254051, 16254052, 16223924, 16223925, 16154521, 16154527, 16253429, 16267611, 16267612, 16243601, 16262281, 16262282, 16253129, 16253130, 16220599, 16223149, 16223922, 16239988, 16248740, 16268536, 16234503, 16234506, 16234508, 16254476, 16254478, 16250930, 16250932, 16253448, 16253617, 16253621, 16254048, 16263201, 16268533, 16198615, 16250938, 16250939, 16250940, 16250969, 16252427, 16252433, 16252434, 16252435, 16252436, 16252437, 16252438, 16252439, 16228920, 16228921, 16259306, 16260131, 16260133, 16260135, 16260136, 16260137, 16261357, 16261358, 16261359, 16210236, 16127365, 16240042, 16245061, 16245068, 16245075, 16220091, 16244765, 16261368, 16244766, 16244768, 16244771, 16261370, 16261374, 16244775, 16221805, 16262290, 16262291, 16262735, 16239351, 16248468, 16244019, 16263236, 16263237, 16263239, 16254465, 16239942, 16239943, 16239944, 16231467, 16259283, 16259285, 16250931, 16250934, 16249663, 16249664, 16262256, 16250423, 16250424, 16250425, 16253115, 16253116, 16245024, 16253117, 16253120, 16243287, 16245027, 16248809, 16248811, 16240418, 16250427, 16250428, 16250430, 16253121, 16240420, 16239950, 16250431, 16262259, 16245042, 16250432, 16241577, 16236705, 16237345, 16270371, 16250422, 16244770, 16254979, 16254980, 16254981, 16254982, 16254983, 16240482, 16240483, 16231451, 16231452, 16252440, 16252441, 16265927, 16265928, 16265929, 16234504, 16234505, 16254481, 16254482, 16248221, 16248239, 16248245, 16130158, 16254461, 16249399, 16253108, 16205679, 16241909, 16205683, 16205685, 16205687, 16203267, 16261355, 16261369, 16127149, 16245359, 16246255, 16246261, 16246263, 16246279, 16246280, 16246284, 16246286, 16263219, 16263220, 16263221, 16263222, 16261380, 16261387, 16261390, 16265972, 16263250, 16263251, 16264487, 16264488, 16264494, 16264495, 16264496, 16264499, 16250933, 16250936, 16268809, 16268810, 16270365, 16270370, 16270374, 16270375, 16130592, 16252453, 16250396, 16250397, 16250398, 16250404, 16253431, 16253433, 16245470, 16245471, 16245473, 16245720, 16245721, 16245722, 16245725, 16245730, 16245772, 16245777, 16231474, 16231475, 16073442, 16267605, 16267606, 16267607, 16248879, 16248883, 16263244, 16263241, 16229190, 16249660, 16262736, 16249439, 16249440, 16249441, 16268808, 16234881, 16234862, 16244002, 16244003, 16244006, 16244020, 16245472, 16245776, 16265966, 16265967, 16265968, 16265970, 16263242, 16263245, 16263246, 16263247, 16268535, 16263248, 16263249, 16264504, 16264506, 16264508, 16265922, 16265925, 16234507, 16239401, 16239404, 16075307, 16274613, 16274615, 16245017, 16227359, 16227360, 16236712, 16239679, 16239686, 16239687, 16248210, 16248212, 16242656, 16254487, 16254538, 16254539, 16254962, 16248302, 16248467, 16269221, 16255631, 16255633, 16255634, 16255636, 16255637, 16266015, 16263202, 16263203, 16263205, 16274614, 16252446, 16252448, 16252451, 16252455, 16252460, 16252466, 16252474, 16266017, 16252477, 16267608, 16250402, 16250403, 16240000, 16248881, 16245723, 16245724, 16245774, 16245775, 16248754, 16248758, 16254458, 16254463, 16254470, 16248457, 16254968, 16254489, 16254960, 16254963, 16242580, 16242581, 16242583, 16241952, 16241958, 16241960, 16242588, 16242589, 16242590, 16242591, 16242592, 16242594, 16242595, 16242598, 16242599, 16242608, 16242610, 16242614, 16242618, 16241986, 16241988, 16241989, 16242555, 16242556, 16242557, 16242624, 16242625, 16081230, 16117958, 16265924, 16265926, 16237331, 16237335, 16237339, 16237342, 16237343, 16250394, 16250395, 16250399, 16250400, 16250401, 16180716, 16240001, 16257567, 16270226, 16231410, 16231417, 16240045, 16240046, 16240047, 16240048, 16240049, 16240050, 16240052, 16240437, 16240438, 16240481, 16239545, 16239632, 16254464, 16269176, 16236703, 16253092, 16270727, 16244323, 16263207, 16263209, 16252481, 16149872, 16261376, 16261377, 16261379, 16261382, 16261383, 16261384, 16261385, 16261388, 16261391, 16261393, 16262242, 16262243, 16262249, 16262253, 16245773, 16239695, 16263255, 16264485, 16264491, 16264492, 16253107, 16253109, 16259296, 16260134, 16260139, 16261363, 16261365, 16261366, 16261367, 16270223, 16270225, 16270227, 16245740, 16245744, 16248760, 16248762, 16248763, 16269937, 16267619, 16244652, 16244653, 16248813, 16248814, 16266022, 16265942, 16262260, 16267625, 16265944, 16265947, 16265950, 16262261, 16262262, 16265956, 16265959, 16265961, 16265962, 16265963, 16250433, 16122583, 16122584, 16259298, 16257566, 16219157, 16192551, 16245741, 16270216, 16270217, 16271690, 16271686, 16223503, 16252380, 16254462, 16254955, 16254959, 16246336, 16273724, 16263233, 16263234, 16263235, 16259301, 16259303, 16245095, 16245097, 16245099, 16245102, 16245104, 16245106, 16245107, 16245108, 16245109, 16245308, 16269912, 16269917, 16269919, 16245743, 16270215, 16254468, 16254469, 16273173, 16273175, 16275387, 16273723, 16273726, 16177166, 16248048, 16228845, 16248471, 16240439, 16240484, 16240485, 16245015, 16254474, 16259287, 16273715, 16262755, 16275716, 16275717, 16275718, 16275720, 16259302, 16254480, 16253114, 16236723, 16236724, 16236725, 16236727, 16236728, 16236729, 16236731, 16236732, 16236733, 16236734, 16236735, 16236737, 16236738, 16236739, 16236741, 16236742, 16236743, 16236744, 16236745, 16236746, 16236747, 16236749, 16236750, 16236751, 16245344, 16259299, 16281682, 16281510, 16281513, 16249414, 16249416, 16249417, 16249418, 16249419, 16249420, 16269914, 16240032, 16240033, 16270222, 16270355, 16248757, 16254061, 16246225, 16246228, 16248458, 16227240, 16265969, 16265971, 16249658, 16268558, 16268565, 16266025, 16266026, 16266027, 16270356, 16270357, 16270359, 16270360, 16032943, 16247524, 16242654, 16236162, 16236163, 16253091, 16253094, 16236706, 16253106, 16145574, 16145575, 16145577, 16145578, 16253110, 16253111, 16253112, 16253113, 16279690, 16268552, 16213235, 16215569, 16205199, 16213248, 16215581, 16215583, 16215584, 16210223, 16210227, 16210231, 16215586, 16268811, 16254073, 16254074, 16254075, 16254076, 16254077, 16254219, 16254242, 16254243, 16254246, 16254448, 16254449, 16254450, 16254452, 16254454, 16273717, 16244319, 16244320, 16239542, 16239544, 16269210, 16269213, 16269217, 16253452, 16226316, 16254046, 16219057, 16219062, 16242616, 16242622, 16241987, 16242559, 16271688, 16221882, 16281511, 16281516, 16271702, 16271703, 16270369, 16261372, 16244773, 16267637, 16267638, 16267639, 16267640, 16267641, 16267643, 16267648, 16268525, 16279651, 16177165, 16262293, 16262294, 16262734, 16262737, 16248044, 16244014, 16245742, 16266071, 16266072, 16266074, 16273225, 16273229, 16273232, 16273234, 16273236, 16273238, 16273241, 16273720, 16240054, 16220575, 16220577, 16220594, 16220595, 16220596, 16246223, 16246226, 16246243, 16273166, 16246345, 16263224, 16234502, 16236704, 16253095, 16253096, 16279724, 16279725, 16279726, 16279727, 16279728, 16279729, 16279730, 16279731, 16279732, 16279733, 16279735, 16282103, 16279737, 16262257, 16262258, 16259307, 16259309, 16267626, 16267627, 16267629, 16275750, 16267631, 16267633, 16267634, 16267636, 16275755, 16244040, 16217215, 16217217, 16217218, 16271689, 16271693, 16237326, 16237327, 16237329, 16237330, 16237333, 16237337, 16237341, 16237344, 16237347, 16237348, 16237349, 16266016, 16252472, 16252484, 16267635, 16268527, 16259281, 16279706, 16279707, 16267610, 16266018, 16266019, 16266020, 16267617, 16267618, 16267620, 16267621, 16267623, 16266021, 16267624, 16269952, 16269954, 16270213, 16270214, 16220584, 16220585, 16220586, 16220587, 16220588, 16220589, 16220590, 16220591, 16220592, 16220593, 16245363, 16270372, 16273728, 16273169, 16254953, 16279693, 16279695, 16229822, 16229981, 16230330, 16236726, 16236740, 16236748, 16244300, 16273150, 16244301, 16281685, 16281686, 16252458, 16219048, 16268538, 16268541, 16220080, 16220081, 16265930, 16265933, 16265938, 16265941, 16265945, 16262292, 16248639, 16229247, 16266013, 16266014, 16266069, 16266070, 16270739, 16270741, 16270743, 16285708, 16270745, 16270747, 16270749, 16270751, 16270753, 16270755, 16270757, 16270759, 16271704, 16271705, 16271706, 16271707, 16271711, 16271713, 16273721, 16258371, 16258376, 16218456, 16240704, 16282110, 16282124, 16280730, 16249445, 16176226, 16271691, 16273154, 16281528, 16270201, 15975033, 16242621, 16242562, 16242623, 16242620, 16230409, 16230414, 16230415, 16230416, 16230417, 16230419, 16230420, 16230421, 16230422, 16230423, 16230425, 16230426, 16230427, 16230430, 16230431, 16230434, 16230436, 16230437, 16230438, 16230439, 16230440, 16230441, 16230443, 16230444, 16230447, 16230450, 16230451, 16230452, 16230453, 16230454, 16230455, 16230458, 16234878, 16234879, 16238704, 16274610, 15734884, 15734886, 15734888, 16290166, 16290167, 16254067, 16254068, 16285867, 16245323, 16176376, 16227362, 16227363, 16227364, 16227365, 16227366, 16227367, 16227368, 16219054, 16219055, 16252483, 16275699, 16291647, 16253098, 16291649, 16291651, 16291652, 16234509, 16234511, 16263227, 16113641, 16254952, 16254957, 16254965, 16254966, 16254069, 16254070, 16254071, 16254072, 16259288, 16211516, 16299155, 16282115, 16282116, 16275698, 16275700, 16275701, 16275702, 16275704, 16275706, 16275708, 16275732, 16048754, 16273181, 16273183, 16273187, 16275721, 16275722, 16275723, 16275725, 16039635, 16039636, 16038548, 16283316, 16283320, 16283321, 16122693, 16280352, 16221121, 16221122, 16221129, 16221130, 16221131, 16221132, 16221133, 16221135, 16282150, 16282151, 16282153, 16253453, 16253620, 16226336, 16254039, 16254044, 16254053, 16254058, 16268542, 16268543, 16268544, 16268545, 16268547, 16268549, 16279688, 16279689, 16279691, 16279723, 16282083, 16282084, 16282085, 16282086, 16282087, 16282088, 16282089, 16282090, 16282091, 16282092, 16282093, 16282094, 16282095, 16282096, 16282097, 16282098, 16282099, 16282100, 16282101, 16282102, 16282104, 16279734, 16279736, 16279738, 16279739, 16282105, 16282106, 16282107, 16282108, 16282109, 16282111, 16282112, 16282113, 16196165, 16095504, 16262254, 16267622, 16262752, 16280731, 16280735, 16280736, 16280737, 16275748, 16295424, 16265940, 16245038, 16250426, 16279743, 16280713, 16279745, 16287651, 16270724, 16265953, 16275752, 16275754, 16262263, 16275766, 16275768, 16265964, 16265965, 16279747, 16273168, 16280716, 16280717, 16295426, 16279748, 16279687, 16279715, 16279716, 16279717, 16279719, 16269944, 16283302, 16283303, 16283305, 16283306, 16283308, 16266023, 16219064, 16243689, 16229985, 16286263, 16271505, 16241955, 16241956, 16113638, 16113639, 16113640, 16294714, 16246339, 16246341, 16246343, 16221273, 16221274, 16221275, 16126055, 16171770, 16171771, 16165876, 16286729, 16283336, 16283337, 16283339, 16283340, 16283341, 16286730, 16286732, 16286733, 16286735, 16283342, 16286737, 16283343, 16283344, 16283345, 16283346, 16286740, 16286742, 16285705, 16279751, 16279752, 16279753, 16279754, 16279755, 16279756, 16279757, 16279758, 16279759, 16279760, 16279761, 16255635, 16255639, 16250937, 16250942, 16245398, 16245399, 16245400, 16245401, 16245403, 16245404, 16245405, 16287622, 16117960, 16117966, 16242600, 16149471, 16286235, 16285863, 16286237, 16286238, 16286240, 16286241, 16285864, 16245406, 16288233, 16245407, 16245408, 16288234, 16237325, 16237346, 16266077, 16236628, 16236629, 16236630, 16236631, 16236632, 16236633, 16236634, 16236635, 16236636, 16236637, 16236674, 16236675, 16236676, 16236677, 16236678, 16236679, 16236680, 16236681, 16236683, 16238708, 16236684, 16236695, 16236696, 16236698, 16236702, 16291658, 16291659, 16294054, 16285848, 16285850, 16285852, 16287634, 16286265, 16292922, 16292924, 16286266, 16286267, 16286410, 16286268, 16286270, 16286271, 16286273, 16286274, 16215564, 16260138, 16243311, 16280718, 16281515, 16281518, 16281519, 16123234, 16149873, 16137389, 16137390, 16245312, 16245314, 16245315, 16245316, 16245317, 16245319, 16280029, 16280030, 16127360, 16275408, 16286748, 16286749, 16286751, 16286753, 16286755, 16286757, 16292927, 16291662, 16223407, 16219972, 16219974, 16219976, 16219978, 16219980, 16219982, 16219983, 16219984, 16219985, 16219986, 16219987, 16219988, 16219989, 16219990, 16219991, 16219992, 16219993, 16219994, 16220065, 16220066, 16220068, 16220069, 16220070, 16252463, 16252469, 16267609, 16220125, 16220126, 16220128, 16220129, 16220130, 16285773, 16220603, 16220607, 16220608, 16220609, 16220610, 16220611, 16220612, 16220613, 16220614, 16296073, 16220092, 16220093, 16213896, 16213898, 16213899, 16267628, 16267630, 16267632, 16201102, 16262283, 16262284, 16254964, 16254969, 16254971, 16254972, 16263223, 16263225, 16254973, 16254974, 16263226, 16263228, 16263229, 16263231, 16262285, 16254976, 16262286, 16262287, 16262288, 16262289, 16196922, 16205097, 16280671, 16273733, 16273735, 16273736, 16273738, 16273739, 16154523, 16154526, 16162667, 16176364, 16239980, 16239981, 16273191, 16273193, 16244764, 16230379, 16230380, 16287639, 16287641, 16248900, 16227137, 16227138, 16227139, 16283327, 16283328, 16270212, 16270228, 16263204, 16263206, 16263208, 16263210, 16263211, 16263212, 16263213, 16283286, 16283287, 16283292, 16283294, 16283295, 16286275, 16286276, 16286277, 16286278, 16286279, 16286280, 16286282, 16286284, 16286286, 16286288, 16286290, 16286293, 16286295, 16279746, 16267613, 16267616, 16290171, 16296064, 16254970, 16292931, 16244697, 16244705, 16244708, 16279749, 16265931, 16265932, 16265934, 16265935, 16265936, 16265937, 16265943, 16265948, 16265951, 16265957, 16265960, 16240490, 16240495, 16240497, 16254975, 16240499, 16254977, 16254978, 16231411, 16275791, 16275792, 16275793, 16275794, 16275795, 16275796, 16275797, 16275798, 16275799, 16275800, 16275801, 16275803, 16275804, 16275805, 16275806, 16275807, 16275808, 16275809, 16275810, 16277358, 16277359, 16277360, 16230410, 16230412, 16230413, 16230424, 16230428, 16230429, 16230435, 16230445, 16230446, 16230448, 16230449, 16230456, 16234880, 16291665, 16291666, 16291668, 16291669, 16291670, 16291671, 16291672, 16291673, 16291675, 16291676, 16296057, 16291987, 16291989, 16291990, 16295395, 16295396, 16295398, 16295399, 16295400, 16295401, 16296058, 16296061, 16295402, 16295403, 16295404, 16295406, 16295407, 16295408, 16295409, 16229243, 16239358, 16239361, 16270803, 16244767, 16244769, 16244774, 16204504, 16286281, 16286283, 16286287, 16286289, 16286291, 16286292, 16286294, 16286296, 16286297, 16286298, 16286299, 16286300, 16286301, 16286302, 16286305, 16286306, 16286307, 16286702, 16286703, 16286705, 16286713, 16286714, 16286715, 16236193, 16236204, 16236205, 16244012, 16244013, 16244016, 16268528, 16268529, 16270720, 16270341, 16270344, 16270346, 16270348, 16270350, 16270722, 16270352, 16270723, 16270218, 16279699, 16280358, 16275712, 16280359, 16270725, 16283332, 16296046, 16280696, 16283333, 16288217, 16296047, 16283335, 16296048, 16288219, 16288218, 16287644, 16285707, 16285709, 16271709, 16285711, 16294046, 16294048, 16294049, 16294051, 16275420, 16275421, 16273727, 16270221, 16241409, 16253436, 16253438, 16221373, 16221374, 16221375, 16221376, 16221377, 16221378, 16221379, 16221380, 16221381, 16221382, 16221383, 16221384, 16213469, 16213470, 16277385, 16283307, 16283329, 16283330, 16282146, 16282148, 16296030, 16296032, 16296034, 16285856, 16285857, 16285844, 16285846, 16285858, 16285860, 16285861, 16285847, 16285849, 16285862, 16285851, 16285853, 16285854, 16285855, 16290972, 16290973, 16290974, 16290975, 16291992, 16291993, 16291994, 16291995, 16291996, 16291998, 16291999, 16292001, 16292003, 16292004, 16292005, 16268530, 16273227, 16254984, 16254985, 16254986, 16254987, 16254988, 16254989, 16255027, 16255028, 16257568, 16257569, 16257570, 16282122, 16282123, 16282125, 16282126, 16282127, 16282128, 16282129, 16282130, 16282131, 16282132, 16282133, 16282134, 16282135, 16282136, 16262266, 16262267, 16262268, 16262269, 16262270, 16262271, 16262272, 16279979, 16279981, 16279983, 16279985, 16279986, 16279971, 16279987, 16280026, 16239535, 16239536, 16239537, 16223146, 16239548, 16239550, 16239552, 16223150, 16239629, 16239630, 16239631, 16239635, 16295425, 16280725, 16295427, 16275703, 16275705, 16266024, 16254460, 16254467, 16217182, 16221499, 16221500, 16221501, 16281563, 16281564, 16281565, 16281566, 16281567, 16281568, 16281569, 16281570, 16281572, 16281573, 16281574, 16282156, 16282157, 16282158, 16282159, 16282160, 16282161, 16282162, 16282163, 16282164, 16282165, 16287627, 16287629, 16287630, 16287631, 16287632, 16287633, 16287635, 16287637, 16287638, 16287640, 16287642, 16287643, 16287645, 16287647, 16290127, 16290128, 16279671, 16279673, 16279675, 16279677, 16279678, 16279679, 16279674, 16279676, 16295418, 16295419, 16295421, 16295423, 16281526, 16281527, 16281529, 16281530, 16281531, 16281533, 16281535, 16281536, 16281538, 16281539, 16281541, 16281542, 16281543, 16281545, 16281546, 16281547, 16281548, 16281549, 16281551, 16245016, 16245018, 16245019, 16245020, 16286728, 16287652, 16287653, 16246224, 16246227, 16219578, 16222556, 16217520, 16215596, 16286706, 16262740, 16263217, 16263218, 16291988, 16291991, 16223921, 16229778, 16296050, 16296051, 16296052, 16287712, 16287706, 16287716, 16290987, 16290988, 16292939, 16292942, 16286745, 16286746, 16285865, 16285866, 16236206, 16295413, 16295414, 16229779, 16283322, 16283323, 16228922, 16228923, 16228924, 16228925, 16288220, 16288221, 16288223, 16288225, 16285782, 16271710, 16271712, 16271714, 16271715, 16271716, 16271717, 16271718, 16271719, 16271720, 16271721, 16273144, 16273145, 16273147, 16273148, 16273149, 16273152, 16273153, 16273155, 16273156, 16234501, 16245322, 16291638, 16291639, 16291641, 16291642, 16282144, 16254064, 16254065, 16285787, 16237328, 16241653, 16297230, 16242593, 16292943, 16292944, 16292947, 16250434, 16250435, 16250436, 16250437, 16271687, 16273174, 16273176, 16286710, 16286712, 16249659, 16245745, 16280679, 16280681, 16280684, 16280687, 16280688, 16280689, 16280690, 16280691, 16280692, 16280693, 16280694, 16271685, 16244631, 16285721, 16269212, 16282117, 16282119, 16275733, 16269215, 16266073, 16245747, 16273160, 16273161, 16273163, 16273165, 16263238, 16285723, 16249391, 16268554, 16280354, 16280355, 16280729, 16287662, 16287663, 16287664, 16287666, 16287668, 16287671, 16287673, 16287676, 16287678, 16287679, 16287681, 16287713, 16287714, 16282140, 16282142, 16290932, 16290933, 16268557, 16268556, 16268562, 16268563, 16268569, 16268573, 16269216, 16288232, 16266075, 16266076, 16253446, 16280373, 16279740, 16279741, 16279742, 16279744, 16286752, 16286754, 16286756, 16286758, 16286761, 16279976, 16279977, 16279978, 16279980, 16279982, 16279984, 16279697, 16266054, 16286744, 16282149, 16270169, 16270200, 16270202, 16270203, 16270204, 16270205, 16270206, 16270207, 16270715, 16279974, 16271698, 16271699, 16269211, 16292946, 16283324, 16270358, 16273206, 16285845, 16288236, 16288238, 16285859, 16288222, 16280363, 16279700, 16286741, 16280028, 16295378, 16280349, 16281517, 16290957, 16292936, 16287717, 16285717, 16279640, 16286734, 16286736, 16286716, 16286717, 16286718, 16286720, 16286721, 16286723, 16286725, 16286727, 16285725, 16285726, 16285727, 16285728, 16287646, 16287648, 16286722, 16286724, 16286726, 16288240, 16295411, 16297228, 16271708, 16220602, 16254484.

Distribution

Worldwide

Quantity

5918 units