FDA Enforcement Class II Terminated

IMRIS ORT100 Table, Part Number 110470-000

Recall: Z-0975-2018 · Reported March 21, 2018

Enforcement

Recall Number
Z-0975-2018
Event ID
79360
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Deerfield Imaging, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 21, 2018
Initiation Date
February 20, 2018
Classification Date
March 13, 2018
Termination Date
August 27, 2019
Address
5101 Shady Oak Rd S, Minnetonka, MN, 55343-4100, United States

Description

IMRIS ORT100 Table, Part Number 110470-000

Reason

The rotational locking mechanisms on the ORT 100/ T2X tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table.

Code Info

Serial Numbers: 10002341, 10003629, 10003801, 10002654, 10002795, 10002796, 10002847, 10003830, 10003243, 10002848, 10003241, 10005651, 10007348, 10007424, 10006886

Distribution

US, Canada, Australia, China, Qatar

Quantity

15 tables