FDA Enforcement
Class II
Ongoing
Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Recall: Z-0974-2024
·
Reported February 14, 2024
Enforcement
- Recall Number
- Z-0974-2024
- Event ID
- 93737
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 14, 2024
- Initiation Date
- January 6, 2022
- Classification Date
- February 2, 2024
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2296, United States
Description
Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Reason
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Code Info
UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030
Distribution
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Quantity
2