FDA Enforcement Class II Ongoing

Cardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2210

Recall: Z-0973-2023 · Reported January 25, 2023

Enforcement

Recall Number
Z-0973-2023
Event ID
91322
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Cardinal Health 200, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 25, 2023
Initiation Date
December 14, 2022
Classification Date
January 18, 2023
Address
3651 Birchwood Dr, N/A, Waukegan, IL, 60085-8337, United States

Description

Cardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2210

Reason

Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.

Code Info

UDI/DI 50885380182479(cs), 20885380182478(ea), Lot Numbers: N210278, N210368, N210423, N210460, N210468 , N210543, N220050, N220099, N220139, N220163

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.

Quantity

0