FDA Enforcement Class II Ongoing

Artis zee biplane MN, Model Number 10094143

Recall: Z-0972-2022 · Reported April 27, 2022

Enforcement

Recall Number
Z-0972-2022
Event ID
89860
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 27, 2022
Initiation Date
February 23, 2022
Classification Date
April 21, 2022
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

Artis zee biplane MN, Model Number 10094143

Reason

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Code Info

UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402

Distribution

US Nationwide.

Quantity

7 systems