FDA Enforcement
Class II
Ongoing
Artis zee floor MN, Model Number 10094142
Recall: Z-0971-2022
·
Reported April 27, 2022
Enforcement
- Recall Number
- Z-0971-2022
- Event ID
- 89860
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 27, 2022
- Initiation Date
- February 23, 2022
- Classification Date
- April 21, 2022
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418, United States
Description
Artis zee floor MN, Model Number 10094142
Reason
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Code Info
UDI: 4056869010090; Serial Numbers: 140421 140115 140101 140205 140213 140204 140209 140219 140224 140218 140226 140221 140220 140206 140105 140423 140406 140422 140102 140214 140109 140202 140424 140114
Distribution
US Nationwide.
Quantity
24 systems