FDA Enforcement Class II Terminated

ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.

Recall: Z-0971-2019 · Reported March 13, 2019

Enforcement

Recall Number
Z-0971-2019
Event ID
81929
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Abbott Laboratories, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 13, 2019
Initiation Date
January 2, 2019
Classification Date
March 1, 2019
Termination Date
June 30, 2021
Address
1921 Hurd Dr, N/A, Irving, TX, 75038-4313, United States

Description

ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.

Reason

Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation.

Code Info

(a) 7D70 21 Lot Number UDI Number 36338UN17 (01)00380740012922(17)190106(10)36338UN17 79865UN17 (01)00380740012922(17)190309(10)79865UN17 12921UN17 (01)00380740012922(17)190503(10)12921UN17 43618UN17 (01)00380740012922(17)190609(10)43618UN17 45337UN17 (01)00380740012922(17)190714(10)45337UN17 84994UN17 (01)00380740012922(17)191012(10)84994UN17 24856UN17 (01)00380740012922(17)191113(10)24856UN17 41652UN18 (01)00380740012922(17)200120(10)41652UN18 99639UN18 (01)00380740012922(17)200226(10)99639UN18 99661UN18 (01)00380740012922(17)200413(10)99661UN18 78493UN18 (01)00380740012922(17)200725(10)78493UN18 95220UN18 (01)00380740012922(17)200907(10)95220UN18 31428UN18 (01)00380740012922(17)201031(10)31428UN18 33784UN18 (01)00380740012922(17)200611(10)33784UN18 (b) 7D70-31 36339UN17 (01)00380740012939(17)190106(10)36339UN17 97389UN17 (01)00380740012939(17)190309(10)97389UN17 12922UN17 (01)00380740012939(17)190503(10)12922UN17 43616UN17 (01)00380740012939(17)190609(10)43616UN17 45338UN17 (01)00380740012939(17)190714(10)45338UN17 84993UN17 (01)00380740012939(17)190901(10)84993UN17 08788UN17 (01)00380740012939(17)191012(10)08788UN17 24857UN17 (01)00380740012939(17)191113(10)24857UN17 41651UN18 (01)00380740012939(17)200129(10)41651UN18 99640UN18 (01)00380740012939(17)200226(10)99640UN18 99662UN18 (01)00380740012939(17)200413(10)99662UN18 33783UN18 (01)00380740012939(17)200509(10)33783UN18 33785UN18 (01)00380740012939(17)200611(10)33785UN18 78859UN18 (01)00380740012939(17)200725(10)78859UN18 95341UN18 (01)00380740012939(17)200907(10)95341UN18 31429UN18 (01)00380740012939(17)201031(10)31429UN18

Distribution

Worldwide Distribution: US (nationwide) and to countries of: CHINA, THAILAND, GUATEMALA, and EL SALVADOR.

Quantity

24,995 units