FDA Enforcement Class II Terminated

Artis zee/zeego, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee/ zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee/ zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Recall: Z-0971-2017 · Reported January 18, 2017

Enforcement

Recall Number
Z-0971-2017
Event ID
76016
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 18, 2017
Initiation Date
December 8, 2016
Classification Date
January 9, 2017
Termination Date
April 9, 2019
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

Artis zee/zeego, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee/ zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee/ zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Reason

Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detection of a fault can result in the failure of a module in the high-voltage generator. - For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with graphics card may, in rare cases, result in the loss of image display in the examination room.

Code Info

Model # 10094135, 10094137, 10094139, 10094141, 10280959, 10848283

Distribution

Nationwide Distribution

Quantity

1,500 distributed Worldwide