FDA Enforcement Class II Terminated

BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy.

Recall: Z-0967-2014 · Reported February 19, 2014

Enforcement

Recall Number
Z-0967-2014
Event ID
67017
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Brainlab AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 19, 2014
Initiation Date
November 15, 2013
Classification Date
February 10, 2014
Termination Date
March 13, 2017
Address
Kapellenstrasse 12, N/A, Feldkirchen, N/A, N/A, Germany

Description

BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy.

Reason

Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect

Code Info

Model/Catalog Code: 55830-25A Instrument Star Unit ML; Serial Numbers (US Devices): 1205112007, 1205112010, 1205112011, 1205112013, 1205112014, 1205112019, 1205112021, 1205112029, 1205112034, 1205112045; Serial Numbers (Non-US Devices): 1205112004, 1205112006, 1205112008, 1205112009, 1205112015, 1205112016, 1205112017, 1205112018, 1205112020, 1205112023, 1205112025, 1205112026, 1205112027, 1205112030, 1205112032

Distribution

Worldwide Distribution - USA including CA, MN, NC, NY, OH, TX and DC. Internationally to Australia, Chile, Germany, Hungary, Italy, Panama, Qatar, Russia, Thailand, United Arab Emirates.

Quantity

25 units