Angled Tip Ureteral Catheter Set, Ureteral Catheter, Global Product No. G14598
Enforcement
- Recall Number
- Z-0965-2020
- Event ID
- 82026
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 12, 2020
- Initiation Date
- January 4, 2019
- Classification Date
- February 4, 2020
- Termination Date
- May 29, 2020
- Address
- 750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States
Description
Angled Tip Ureteral Catheter Set, Ureteral Catheter, Global Product No. G14598
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
8023195 8314384 8409525 NS8040177 7857489
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.