FDA Enforcement Class II Terminated

PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10

Recall: Z-0962-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-0962-2020
Event ID
82026
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2020
Initiation Date
January 4, 2019
Classification Date
February 4, 2020
Termination Date
May 29, 2020
Address
750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States

Description

PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10

Reason

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Code Info

7832498 7832499 7832500 7925385 7925387 7930517 8090868 8287793 8287794 8287799 8400121 8548813 8548815 NS8548814 7760235 7881599 8062433 8074027 8290383 8290384 8404559 8404562 7999663 8077582 8077583 8065959 8073949 8074014 8074015

Distribution

Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.

Quantity

3929