FDA Enforcement Class II Completed

4.0mm x 12mm Cannulated Screw, Short Thread; Part number 3005-40012

Recall: Z-0959-2022 · Reported April 27, 2022

Enforcement

Recall Number
Z-0959-2022
Event ID
89771
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Acumed LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 27, 2022
Initiation Date
March 9, 2022
Classification Date
April 20, 2022
Address
5885 NW Cornelius Pass Rd, Hillsboro, OR, 97124-9370, United States

Description

4.0mm x 12mm Cannulated Screw, Short Thread; Part number 3005-40012

Reason

Cannulated Screws 4.0mm x 10mm and 4.0mm x 12mm may not have sufficient axial pullout strength for the indicated use. Remote possibility of resulting in a mal or nonunion of the bone fragments, iatrogenic fracture, joint impingement or damage, or soft tissue damage.

Code Info

All Batch numbers distributed between 22 AUG 2018 to 07 FEB 2022. UDI:10806378022876

Distribution

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, Puerto Rico and the country of South Africa.

Quantity

506 unit