FDA Enforcement Class I Ongoing

GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR

Recall: Z-0958-2024 · Reported February 21, 2024

Enforcement

Recall Number
Z-0958-2024
Event ID
93780
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Wipro GE Healthcare Private Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 21, 2024
Initiation Date
December 26, 2023
Classification Date
February 9, 2024
Address
122 Part 1, Odyssey And Galileo, Bengaluru, N/A, N/A, India

Description

GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR

Reason

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Code Info

a) Lullaby Incubator, Model Numbers: M1151042, QCR031; b) Lullaby Incubator XP, Model Numbers: 8320, M1151135, QCR034; c) Lullaby Incubator TR, Model Numbers: M1151357 ALL SERIAL NUMBERS

Distribution

Worldwide distribution.

Quantity

1410 units