GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000
Enforcement
- Recall Number
- Z-0957-2024
- Event ID
- 93780
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Wipro GE Healthcare Private Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 21, 2024
- Initiation Date
- December 26, 2023
- Classification Date
- February 9, 2024
- Address
- 122 Part 1, Odyssey And Galileo, Bengaluru, N/A, N/A, India
Description
GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus 3000, e) Care Plus 4000, Plus 4000
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
a) Care Plus, Model Numbers: 6600-0040-900, 6600-0203-900, 6600-0203-903, 6600-0232-910, 6600-0232-920, 6600-0235-901, CARE-PLUS; b) Care Plus 1000, Model Numbers: 8200, 6600-0264-901, 6600-0264-903, 6600-0265-901, 6600-0265-902, 6600-0265-903, 6600-0265-908, 6600-0266-901, 6600-0266-903, 6600-0267-901, 6600-0269-901, 6600-0269-903, 6600-0270-901, 6600-0272-901; c) Care Plus 2000, Model Numbers: 2073175-001, 6600-0273-901, 6600-0274-901, 6600-0274-903, 6600-0274-911, 6600-0275-901, 6600-0275-902, 6600-0275-903, 6600-0275-908, 6600-0275-911, 6600-0276-901, 6600-0276-903, 6600-0277-901, 6600-0279-903, 6600-0280-901, 6600-0280-903, 6600-0281-903, 6600-0282-901, M1050218, M1160968, M1206249, M1242714, QCR043; d) Care Plus 3000, Model Numbers: 6600-0203-902, 6600-0223-981, 6600-0230-900, 6600-0230-901, 6600-0232-901, 6600-0232-911, 6600-0232-921, 6600-0232-980, 6600-0232-983, 6600-0233-900, 6600-0233-901, 6600-0233-911, 6600-0235-900, 6600-0293-901, 6600-0293-903, 6600-0294-901, 6600-0294-902, 6600-0294-903, 6600-0294-905, 6600-0294-908, 6600-0294-909, 6600-0295-901, 6600-0301-901, 6600-0301-903, 6600-0302-901, 6600-0302-903, 6600-0303-901, 6600-0333-903, 6600-0463-901, 6600-0511-901, 6600-0514-900, 6623-0010-903, 6623-0010-923, 6623-0011-902, 6623-0011-903, 6623-0011-920, 6623-0011-923, 6623-0021-920, 6623-0111-900, 6623-0111-903, 6623-0111-905, 6623-0311-900, 6623-0311-903, 6623-0411-900, 6623-0411-905, 6623-0511-900, 6623-0511-902, 6623-0511-903, 6623-0511-905, 6623-0511-910, 6623-0511-913, 6623-0511-920, 6623-0511-921, 6623-0511-943, 6623-0521-900, 6623-0521-902, 6623-0521-905, 6623-0521-910, 6623-0521-943, 6623-0610-905, 6623-0611-900, 6623-0611-903, 6623-0611-905, 6623-0611-943, 6623-0621-943, 6623-0711-900, 6623-0711-903, 6623-0711-920, 6623-0811-900, 6623-0811-903, 6623-0811-910, 6623-0811-913, 6623-0811-920, 6623-0811-943, 6623-0911-900, 6623-0911-905, 6623-0921-905, 6623-1011-903, 6623-1411-900, 6623-1421-905, 6623-1511-900, 6626-0010-903, 6626-0010-913, 6626-0010-923, 6626-0011-900, 6626-0011-902, 6626-0011-903, 6626-0011-905, 6626-0011-920, 6626-0011-923, 6626-0021-903, 6626-0111-900, 6626-0121-905, 6626-0121-920, 6626-0211-900, 6626-0221-900, 6626-0221-903, 6626-0311-900, 6626-0311-902, 6626-0311-903, 6626-0311-905, 6626-0311-910, 6626-0321-900, 6626-0411-900, 6626-0411-903, 6626-0411-905, 6626-0411-920, 6626-0421-900, 6626-0421-913, 6626-0421-920, 6626-0510-900, 6626-0510-905, 6626-0511-900, 6626-0511-901, 6626-0511-902, 6626-0511-903, 6626-0511-905, 6626-0511-910, 6626-0511-913, 6626-0511-920, 6626-0511-922, 6626-0511-923, 6626-0521-900, 6626-0521-902, 6626-0521-903, 6626-0521-905, 6626-0521-913, 6626-0521-920, 6626-0521-923, 6626-0610-923, 6626-0611-900, 6626-0611-903, 6626-0611-905, 6626-0611-920, 6626-0611-923, 6626-0621-903, 6626-0710-903, 6626-0711-900, 6626-0711-903, 6626-0711-905, 6626-0810-923, 6626-0811-900, 6626-0811-902, 6626-0811-903, 6626-0811-904, 6626-0811-905, 6626-0811-920, 6626-0811-923, 6626-0821-900, 6626-0821-903, 6626-0821-920, 6626-0911-900, 6626-0911-902, 6626-0911-903, 6626-0911-905, 6626-0911-923, 6626-0921-902, 6626-0921-903, 6626-0921-905, 6626-1011-900, 6626-1111-905, 6626-1421-905, 6626-1511-900, 6626-1521-900, 6626-1611-920, 6626-1811-900, 6628-0011-900, 6628-0011-903, 6628-0011-913, 6628-0021-903, 6628-0111-900, 6628-0111-903, 6628-0211-900, 6628-0221-900, 6628-0311-900, 6628-0321-900, 6628-0321-910, 6628-0411-900, 6628-0411-902, 6628-0411-903, 6628-0421-900, 6628-0511-900, 6628-0511-920, 6628-0511-923, 6628-0511-943, 6628-0521-943, 6628-0610-923, 6628-0611-900, 6628-0611-903, 6628-0611-905, 6628-0621-900, 6628-0711-905, 6628-0721-905, 6628-0810-923, 6628-0811-900, 6628-0811-902, 6628-0811-903, 6628-0811-905, 6628-0811-910, 6628-0811-920, 6628-0811-922, 6628-0820-923, 6628-0821-900, 6628-0821-903, 6628-0910-900, 6628-0911-900, 6628-0911-903, 6628-0911-905, 6628-0921-902, 6628-0921-920, 6628-1321-920, 6628-1421-903, 6628-1521-920, M1050214, M1051981, M1052131, M1052601, M1053655, M1053786, M1054018, M1056216, M1056217, M1061559, M1117929, M1136797, M1137588, M1137746, M1137750, QCR045; e) Care Plus 4000, Plus 4000, Model Numbers: 6600-0203-901, 6600-0225-903, 6600-0232-900, 6600-0325-901, 6600-0325-903, 6600-0326-901, 6600-0326-902, 6600-0326-903, 6600-0326-904, 6600-0326-909, 6600-0327-901, 6600-0333-901, 6600-0333-902, 6600-0334-901, 6600-0334-902, 6600-0334-903, 6600-0334-904, 6600-0334-905, 6600-0334-910, 6600-0334-911, 6600-0421-900, 6624-0111-900, 6624-0111-903, 6624-0311-900, 6624-0311-903, 6624-0321-900, 6624-0321-903, 6624-0321-943, 6624-0321-967, 6624-0411-900, 6624-0421-900, 6624-0421-905, 6624-0511-900, 6624-0511-903, 6624-0511-913, 6624-0521-900, 6624-0521-903, 6624-0611-903, 6624-0621-902, 6624-0811-900, 6624-0821-920, 6624-0821-943, 6624-0911-900, 6624-0921-900, 6624-1021-900, 6624-1311-900, 6624-1321-900, 6624-1321-903, 6624-1421-900, 6624-1421-903, 6624-1511-900, 6624-1521-900, 6627-0010-913, 6627-0011-903, 6627-0020-913, 6627-0021-900, 6627-0021-903, 6627-0021-913, 6627-0021-923, 6627-0110-923, 6627-0111-900, 6627-0111-905, 6627-0111-920, 6627-0121-900, 6627-0121-903, 6627-0121-920, 6627-0211-900, 6627-0221-900, 6627-0221-903, 6627-0221-905, 6627-0221-910, 6627-0311-900, 6627-0311-903, 6627-0311-905, 6627-0320-903, 6627-0320-913, 6627-0321-900, 6627-0321-902, 6627-0321-903, 6627-0321-910, 6627-0321-923, 6627-0321-966, 6627-0321-967, 6627-0411-900, 6627-0411-903, 6627-0421-900, 6627-0421-902, 6627-0421-903, 6627-0421-905, 6627-0421-910, 6627-0421-913, 6627-0421-920, 6627-0421-966, 6627-0421-967, 6627-0511-900, 6627-0511-903, 6627-0511-920, 6627-0511-923, 6627-0521-900, 6627-0521-903, 6627-0521-913, 6627-0521-920, 6627-0521-923, 6627-0610-923, 6627-0611-900, 6627-0611-903, 6627-0621-903, 6627-0711-900, 6627-0811-900, 6627-0811-903, 6627-0811-905, 6627-0820-913, 6627-0821-900, 6627-0821-923, 6627-0911-900, 6627-0911-903, 6627-0911-905, 6627-0911-910, 6627-0921-900, 6627-0921-903, 6627-1011-903, 6627-1021-903, 6627-1121-922, 6627-1311-900, 6627-1321-900, 6627-1321-903, 6627-1321-910, 6627-1321-966, 6627-1411-900, 6627-1411-920, 6627-1421-900, 6627-1421-903, 6627-1421-905, 6627-1421-910, 6627-1421-920, 6627-1421-953, 6627-1511-900, 6627-1921-900, 6629-0011-900, 6629-0011-905, 6629-0021-900, 6629-0021-903, 6629-0021-943, 6629-0110-923, 6629-0111-900, 6629-0121-900, 6629-0121-910, 6629-0121-923, 6629-0211-900, 6629-0221-900, 6629-0221-910, 6629-0310-923, 6629-0311-900, 6629-0311-903, 6629-0320-913, 6629-0320-923, 6629-0321-900, 6629-0321-903, 6629-0321-910, 6629-0321-913, 6629-0321-920, 6629-0321-923, 6629-0411-900, 6629-0411-903, 6629-0411-905, 6629-0420-913, 6629-0421-900, 6629-0421-902, 6629-0421-903, 6629-0421-910, 6629-0421-913, 6629-0511-900, 6629-0511-903, 6629-0521-900, 6629-0521-902, 6629-0610-923, 6629-0611-900, 6629-0611-910, 6629-0620-923, 6629-0711-900, 6629-0711-910, 6629-0810-923, 6629-0811-900, 6629-0811-902, 6629-0811-903, 6629-0811-923, 6629-0820-923, 6629-0821-900, 6629-0821-902, 6629-0821-903, 6629-0821-920, 6629-0821-923, 6629-0911-900, 6629-0920-903, 6629-0920-923, 6629-0921-900, 6629-0921-902, 6629-0921-905, 6629-1311-900, 6629-1321-900, 6629-1321-903, 6629-1411-900, 6629-1421-900, 6629-1421-903, 6629-1421-920, 6629-1511-900, M1052135, M1052600, M1053785, M1137897; ALL SERIAL NUMBERS
Worldwide distribution.
21626 units