BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.
Enforcement
- Recall Number
- Z-0954-2021
- Event ID
- 87178
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 17, 2021
- Initiation Date
- January 7, 2021
- Classification Date
- February 5, 2021
- Termination Date
- December 3, 2021
- Address
- 7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States
Description
BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.
Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may be mislabeled. This could result in a kit containing two pouches of N1 master mix or two pouches of N2 master mix, instead of one pouch of each. If laboratory personnel do not notice that both pouches contain tubes with the same color foil seal and use two of the same tubes, there is a low potential for a false negative SARS-CoV-2 result. A false negative SARS-CoV-2 result could lead to an increased risk of disease progression for the patient and additional spread of SARS-CoV-2 if the patient is not quarantined.
Catalog Number: 444213, Lot Numbers: K20-507, K20-508, K20-509, K20-516, K20-517, K21-020, K21-021. GTIN Number: 00382904442132
US nationwide distribution.
938 kits