FDA Enforcement
Class II
Terminated
Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained healthcare professionals and clinicians to access archived multimedia data, forms and patient information.
Recall: Z-0952-2021
·
Reported February 10, 2021
Enforcement
- Recall Number
- Z-0952-2021
- Event ID
- 87086
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Topcon Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 10, 2021
- Initiation Date
- December 18, 2020
- Classification Date
- February 4, 2021
- Termination Date
- October 19, 2022
- Address
- 111 Bauer Dr, Oakland, NJ, 07436-3123, United States
Description
Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained healthcare professionals and clinicians to access archived multimedia data, forms and patient information.
Reason
Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record
Code Info
Software Version 3.1
Distribution
AZ, MO, NC, NY, PA, TX
Quantity
13 units